HT and single tank error diagnosis
6 - 9
Siemens-Elema AB
SPR8-215.898.02
Page 9 of 12
MOBILETT Plus/ Plus E
Solna, Sweden
Rev. 01
01.02
SPS-UD
Exposure values, 133 kV 5 mAs, HTPOS & HTNEG 2 V/div,
20 kV/div; 200 µs/div (only the end of the exposure is shown)
Fig. 6
HT-signal, Arcing between plus and minus side (Defective tube). Replace single tank.
HTPOS on
D1
HTNEG on
D1