Display messages
2 - 17
Siemens-Elema AB
SPR8-215.898.02
Page 17 of 22
MOBILETT Plus/ Plus E
Solna, Sweden
Rev. 01
01.02
SPS-UD
Err 32
Service mode
Test program
P03
The capacitance of
the capacitor bank is
too low
•
One or more capacitor rows are
disconnected. If there is an internal
error in a capacitor, the whole row
is disconnected by a fuse, refer to
page 130** and section "Capacitor
bank".
•
If the error appears after working
on the capacitor bank, it may be
that the internal residual voltage in
the modules was too high when
they were connected to each other.
This caused the fuse to blow, refer
to page 130**.
Err 33
The mes-
sage is not
shown to
the opera-
tor.
Generator ON
Temporary signal
fault during the pro-
gram execution.
•
"Err 33" messages can be ignored.
This message can only be read out
from the message history, P08.
The message is only used as aid
for the software development.
•
Acknowledge and clear the mes-
sage history.
Err 34
The mes-
sage is not
shown to
the opera-
tor. This
message
can only
be read
out from
the mes-
sage his-
tory, P08.
Generator ON
The number of mes-
sages is beyond the
maximum limit.
Messages are being lost and not dis-
played for the customer. All mes-
sages are stored in the error
memory.
•
Using section "P08 - Display of
message protocol ", display and
record the messages contained in
the error memory. Using the error
message list, try to determine
which module, board, etc. is
generating the messages.
Dis-
played
messages
Mode of
operation
Cause
Possible reasons and advice
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1