Siemens AG
TD00-000.841.02.04.02
Display
05.05
CSPS21
System-related Adjustments
11
Page 11 of 22
Medical Solutions
Size and Position
Check the size and position in the norm being used. The picture must be displayed “full
screen".
If this is not the case, if needed, the size and position (only position is possible) must be
performed manually.
Frequency / Phase (Focus)
NOTE
This adjustment ensures that the picture is correct for the screen
size.
Incorrect adjustment is manifested as unsharp focus.
The adjustment can be performed using the SMPTE test pattern
(quick adjustment) or by generating a text page.
•
Using SMPTE:
Tab. 3
SMPTE
•
Using the Text Page (more exactly adjustable)
Select Explorer to specify the background color of Notepad and to select Notepad.
Select Explorer:
- Use the “Windows and E” buttons to open Explorer.
- Open “Control Panel” in Explorer.
Select Display
(invert so that the highlighted elements appear on a black background):
- Open “ Display” => select “ Appearance” => under Items, select “ Selected Items” =>
under Color, select “ black” => accept with apply.
Generate a text page with “mmmm”:
Select the SMPTE test pattern in the
system.
The 5 high-frequency line groups must
be equally well visible!
The adjustment is performed under: =>
Menu => Other => using the sliders:
Frequency and Phase
Fig. 5:
SMPTE Leonardo