5. Performing troubleshooting and maintenance
Warning messages
CODE MESSAGE
PROBABLE CAUSE
POSSIBLE SOLUTION
01-13
Test strip or LQC vial not valid
Stride encountered an unknown error
scanning the test strip or vial.
Scan the test strip or vial again. If the error continues,
cancel the test and repeat using a new test strip from
the vial.
01-16
Data integrity error
Stride encountered an unknown
internal error.
Turn Stride on and off.
Error messages
CODE MESSAGE
PROBABLE CAUSE
POSSIBLE SOLUTION
02-01
Apply sample timeout
You didn't apply the sample within 2 minutes.
Redo the test and apply the sample when
prompted by Stride.
02-02
Used test strip inserted
You attempted to use an already processed test
strip.
Discard test strip and restart test with a new strip.
02-03
Early fill error
You attempted to apply a patient or QC sample
before Stride was ready.
Redo the test and only apply the sample when
prompted by Stride.
02-04
Early test strip ejection
You ejected the test strip before Stride was finished
processing the results.
Discard test strip and restart test with a new strip.
02-05
Expired test strip
The test strip is beyond the expiration date.
Verify the Stride date setting is correct and then
redo the test using a test strip from a new vial.
02-07
Test strip port protective cap
removed
You removed the test strip port protective cap
during a test.
Align tab in cap with slot and slide cap until it clicks
into place.
Redo the test with a new strip.
02-08
Test aborted
You canceled a test in progress.
Redo the test with a new strip.
02-09
Critical battery level
Your batteries are almost out of power.
Replace with 3 new AA batteries.
02-10
Communications error
Stride encountered an unknown
communication error.
If you are connected to a PC, disconnect the USB
cable and then reconnect it again to the PC.
If you are upgrading the Stride firmware, cancel
and restart the upgrade.
02-11
Heater control error
Stride encountered an error with the internal heater.
Turn Stride off and on. If the error persists, you may
need to replace the Stride analyzer.
Содержание XPRECIA STRIDE
Страница 2: ...XPRECIA STRIDE USER GUIDE ...
Страница 4: ...1 Start here ...
Страница 6: ...2 1 Exploring Stride ...
Страница 20: ...30 2 Per forming a patient test ...
Страница 26: ...42 3 Per forming a LQC test ...
Страница 30: ...50 4 Ch anging the settings ...
Страница 35: ...60 5 Performing troubleshooting and maintenance ...
Страница 43: ...76 Appe ndix A Specifications ...
Страница 45: ...80 Appendix B Safety ...
Страница 48: ...86 A ppendix C Operational theory ...
Страница 51: ...92 Appendix D Symbols ...
Страница 53: ...96 Appendix E Glossary ...