53
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
OPERATION
4
4.12
Checking the analog output
4.12.1
Set test current for analog output (command ITEST)
Starts and stops the mode for checking the current output.
Syntax:
ITEST [Current]<cr>
Example:
4.13
Relay operation
4.13.1
Display/set relay operating mode (command RELAY_MODE)
Sets the relay operating mode,
see “Display/set relay operating mode (command RELAY_-
.
Syntax:
RELAY_MODE
[warn-alarm / fault_alarm / high_open / low_open]<cr>
4.13.2
Display/set relay switching point (command RSEL)
Sets the relay switching points.
Syntax:
RSEL
<cr>
Example:
Where
current = test current (mA)
>itest 4
Test current set to 4 mA. Use ITEST to stop Test mode.
>itest
Current Test mode stopped.
>
warn_alarm
= Relay open when maintenance request exists for error
fault_alarm
= Relay open when error exists
high_open
= Relay open when measuring result above upper point
Relay closed when measuring result below lower point
low_open
= Relay open when measuring result below lower point
Relay closed when measuring result above upper point
NOTICE:
The contact relay is momentary.
>rsel
LO POINT (%02) : 10.0 ?
HI POINT (%02) : 11.0 ?