39
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
OPERATION
4
4.5.8
Scaling the analog output (ASCL)
The analog output can be scaled freely.
1
After pressing
Ent
, use submenu item
Set Lo
to set the oxygen value to be transferred
with the lower mA value (4 mA or 0 mA).
2
Use submenu item
Set Hi
to set the oxygen value to be transferred with the upper mA
value (20 mA).
Fig. 28: Scaling the analog output
4.5.9
Digital output (ALA)
1
Press
Ent
to display the current switching position.
2
To check the switching function, press
Ent
and use
Dn
and
Up
to select the desired
switching function
OPE
(open) or
CLO
(closed).
Fig. 29: Checking and changing the digital output state
4.5.10
Resetting the measuring device (rESE)
The device is restarted.
Fig. 30: Restarting the oxygen sensor TRANSIC111LP
Ent
Ent
Dn
Back
Back
k
Up
Ent
Ent
Ent
Ent
Dn
<
<
<
<
<
<
<
<
<
<
Up
<
<
Back
Back
k
3 x
6 x
2 x
Ent
Ent
Ent
Ent
<
<
<
Dn
Up
Ent
Ent
Ent
Ent
Ent
Ent
<
<
<
<
Back
Back
k