Parameter
Type
Description
Visibility
Access
DeviceRegistersValid
bool
Returns if the current register set is valid and consis‐
tent.
Updated after the
DeviceRegistersStreamingEnd
com‐
mand.
Expert
R
DeviceTemperature[DeviceTem‐
peratureSelector]
float
Device temperature in degrees Celsius (C).
Beginner
R
Table 9: Device control custom parameters
Parameter
Type
Description
Visibility
Access
DeviceRegistersStreamingActive
bool
True if a registers streaming session is active.
Guru
R
DeviceTemperatureMin
int
The minimum temperature ever in degrees Celsius (C)
measured at the location selected by DeviceTempera‐
tureSelector.
Beginner
RO
DeviceTemperatureMax
int
The maximum temperature ever in degrees Celsius (C)
measured at the location selected by DeviceTempera‐
tureSelector.
Beginner
RO
DeviceCurrentBootPath
enum
The booted software path, i.e., bootloader, kernel, appli‐
cation etc. Primary is the normal path and Secondary is
used for rescue purposes if the primary path fails.
Values:
Primary
Secondary
Expert
R
DeviceNextBootPath
enum
The boot path that will be used after the next soft
reset
1
. This is normally set to secondary to boot the res‐
cue version of the system if an unexpected reset
occurs.
Values:
Unknown
Primary
Secondary
Guru
R
DeviceBootCount
int
Accumulated number of times the device has been
booted.
Beginner
RO
DeviceOperationTime
int
Accumulated operation time of the device, in seconds.
Beginner
RO
1
A soft reset is an unwanted re-boot of the device.
Table 10: Device control standard SFNC commands
Command
Description
Visibility
Access
DeviceReset
Resets the device to its power up state.
Guru
RW
DeviceRegistersStreamingStart
Prepare the device for registers streaming without
checking for consistency.
DeviceRegistersStreaming is used when loading and
saving a device's configuration to/from file on the host.
Guru
W
DeviceRegistersStreamingEnd
Announce the end of registers streaming.
After this command
DeviceRegistersValid
is updated to
reflect the status of the device settings.
Guru
W
CONFIGURATION
7
8020774/14IM/2019-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | Ranger3
37
Subject to change without notice