11.2.5
Measuring head 10 mm
1
2
3
59
(2.32)
75.5
(2.97)
Ø
4
0
(1
.5
7
)
68.7
(2.70)
10
(0.39)
2
8
°
Figure 19: structure and device dimensions, unit: mm (inch), decimal separator: period
1
Optical fiber connection
2
Average measuring distance (typical)
3
Measuring range
11.3
Performance
Measurement procedure
Chromatic confocal
Measured values
•
Distance measurement
•
Coating thickness measurement
Measuring rate
Max. 10 kHz
Chromatic confocal meas‐
uring range
Depending on the measuring head used
Gradient error
1)
< 10
-3
Linearity deviation
1)
< 3,3 x 10
-4
x upper measuring range limit
Axial resolution
3 x 10
-6
x upper measuring range limit
Number of measuring
channels
1
Light sender
LED, visible white light
LED risk group
RG1 (IEC 62471)
1)
Measurement accuracy = linearity dev (slope error x measured value). The specified accuracy is
achieved at room temperature.
11.4
Interfaces
OD7000 Core
OD7000 Prime
Ethernet
TCP/IP
Data transmission rate: 100 Mbit/s
10BASE-T/ 100BASE-TX
TECHNICAL DATA
11
8027188//2022-05-25 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD7000
59