Setting
Meaning
Hold function ON/reset
If the hold function is activated, the external input terminal is used
as a hold input or reset input.
Peak Hold/Bottom Sample Hold: Hold input
Auto Peak Hold/Auto Bottom Hold: Reset input
Start process recording
Used as a recording input.
If this setting is activated, recording is performed. When the speci‐
fied amount of saved entries is reached, the recording is stopped
and a file is created.
If the input setting is set to
OFF
during the recording, the process is
paused and a file is created using the data saved up until that
point.
Measured value offset
Used as an offset or offset release input to shift the zero point.
< 1 s: Offset
≥
1 s: Reset offset
Teach-in
The following switching functions (teach-in of limits) are performed
depending on the duration of the signal at the MF input:
Pulse duration (ms)
Function
0 - 100
No operation
100 - 199
Ch1 Upper limit teach
200 - 299
Ch1 Lower limit teach
300 - 399
Ch2 Upper limit teach
400 - 499
Ch2 Lower limit teach
500 - 599
Ch3 Upper limit teach
600 - 699
Ch3 Lower limit teach
700 - 799
Ch4 Upper limit teach
800 - 899
Ch4 Lower limit teach
> 900
No operation
Recording of single values
With every trigger signal, a new measured value is saved and
stored in a file on the device. Select the
Start recording
function to
create a new file. Select the
Stop recording
function to close and
save this file,
see "Starting and testing the recording", page 59
The default mode is
None
.
7.3.9.3
Setting the debounce
Sets the delay time with which the external input is switched on until actual operation
begins. This prevents the external input from suddenly being switched on or off.
Values are possible from 0 ms to 32,767 ms, the standard value is 20,000 ms.