13.2.5
Power supply
Table 17: Voltage supply
Voltage supply
Power input
•
Analyzer
•
Heated sample gas line
•
Gas sampling unit
•
Heated probe tube
Power input
•
Approx. 1000 VA
•
Approx. 95 VA/m
•
Approx. 450 VA
•
Approx. 450 VA
Table 18: Optional interfaces
Interfaces (optional)
Digital outputs
4 outputs, 24 V, 0.5 A
Digital inputs
Electrically isolated, 24 V, 0.3 A
13.2.6
Gas supply
Important information
NOTICE
Malfunction of the measuring device due to unsuitable instrument air
Operation with air not satisfying the specifications voids the warranty and does not
ensure proper functioning of the measuring device.
b
Only feed conditioned instrument air to the measuring device.
b
The instrument air quality must meet the specification.
Supply gases
Table 19: Supply gases
Gas
Quality
Inlet pressure
Flow rate
Instrument air (zero
gas quality)
Particle size max. 5
μ
m
Pressure dew point max. –40
°C
Oil content max. 0.01 mg/m
3
ISO 8573-1:2010 [1:4:2]
600 ... 700 kPa
(6.0 ... 7.0 bar)
Approx. 350 l/h
Approx.. 1300 l/h
(with backflush)
Inlet of instrument
air solely as induc‐
tion air for ejector
Particle size max. 5
μ
m
Pressure dew point max. +3
°C
Oil content max. 0.1 mg/m
3
ISO 8573-1:2010 [1:4:2]
500 ... 700 kPa
(5.0 ... 7.0 bar)
Approx. 1300 l/h
Air dryer FRL-00025
(option)
With the air dryer option,
approx. 3,000 l/h additional
instrument air are required
External span gas
Span gas must comply with
the specifications of the
standards to be applied
Max. 400 kPa
(3.5 bar)
Approx. 350 l/h
13.2.7
Tube connections
Table 20: Tube connections
Connection
Dimension
Sample gas inlet
Clamping ring screw connection (hose fitting)
4 mm inner diameter
6 mm outer diameter
TECHNICAL DATA
13
8026608/AE00/V1-0/2021-10 | SICK
O P E R A T I N G I N S T R U C T I O N S | MCS200HW-MP
57
Subject to change without notice