Operating instructions
Chapter
3
ICR 84x Image Code Reader
Product description
8 010 961/0000/10-05-2005
©
SICK AG · Division Auto Ident · Germany · All rights reserved
3-3
11. For configuring the ICR via the auxiliary interface a 3-core RS 232 data cable (null
modem cable) with two 9-pin D-Sub sockets for connecting the PC to the auxiliary
interface of the ICR in the CDB 420 or CDM 420 Connection Module,
e. g. no. 2 014 054. Pin 2 (RxD) and pin 3 (TxD) are crossed.
12. For configuring the ICR and outputting images via the Ethernet interface:
For the connection of the ICR to the network: a standard Ethernet data cable (patch
cable), e.g. no. 6 026 083.
For the connection of the ICR to the PC (network card): an Ethernet crossover line,
e. g. no. 6 026 084.
13. To use the online help system “CLV Setup Help“, an HTML browser is required,
e. g. “Internet Explorer
TM
“ browser.
3.1.4
Product features and functions (overview)
High-performance scanner:
•
CMOS sensor of newest technology (1.3 Mega pixel)
•
Fixed focus
•
Front reading window, variant: side reading window
•
Reading range (DOF) 58 to 106 mm (2.29 to 4.18 in), standard type
•
Reading area (dependent of the reading distance), 40 mm x 32 mm (1.58 in x 1.26 in),
at reading distance 80 mm (3.15 in), standard type
•
Resolution 0.15 to 2.0 mm (5.9 to 78.7 mil), standard type
•
Image refresh rate/decoding frequency 25 Hz at highest resolution
•
Function to adjust on the code print quality
•
Variable active evaluation range of the image buffer, thereby higher image refresh rate
Safety and user-friendly features:
•
Robust, compact metal housing, max. IP 65, CE certification
•
LED class 1, illumination diodes switches off if reading interval is active for too long and
if the output power is exceeded
•
Automatic self-test on startup. Can also be triggered at any time
•
Diagnosis tools for installing and monitoring the system
•
Parameterized output of reading diagnosis data in the read result
•
Operating data query, and error messages on request
•
Test string function for signaling readiness for operation
•
Future proof thanks to firmware update via serial data interface (flash PROM)
•
Low power consumption, other voltage range
Easy operation/configuration:
•
With "CLV Setup" PC software for Windows (online), Help system and four assistants
(Connection Assistant, Ethernet Assistant, Scanner Adjustment, and DataMatrix-
AutoSetup).
Displaying the image memory content via the ImageFTP program.
•
Alternatively with simple command strings, also for use with special devices
•
Four status LEDs
•
Beeper to confirm device functions or operating steps (can be switches off)