6.2.2.1
Wiring of the temperature sensor
GND
CAN
Power
L
H
GND
+24
Analog In
W
H
B
R
Figure 14: Wiring connections on the control unit
1
Connection line, temperature sensor
NOTE
After connecting, the "External" switch in the "Temperature Input" menu must be set to
"Yes" (
see "Setting the use of an external temperature sensor", page 60
.
6.3
Uninstalling the electrical system
DANGER
Electrical accident through bare, live lines
When uninstalling, unsecured, live lines can lead to serious accidents.
b
Switch the power supply to the device off before starting uninstallation.
b
If power supply is required during uninstallation: Secure all live lines during unin‐
stallation work so that nobody can be injured.
CAUTION
Electrical accident through unsecured switches
Switches that should no longer be switched on for safety reasons can lead to serious
accidents when switched on accidentally.
b
Replace defective switches.
b
As long as defective switches have not been replaced: Secure against being
switched on accidentally with appropriate signs or safety locks.
NOTICE
Device damage through incorrect storage of electrical cables
Incorrect storage can lead to cable ends becoming damp and soiled. This can lead to
device damage when restarting the device.
b
Insulate cable ends.
b
Protect cable ends against dirt and dampness.
6
ELECTRICAL INSTALLATION
32