Error message
Description
Possible cause/clearance
DSP: INV PARA
Incompatible software (SR-unit)
Invalid values entered.
b
Check software version; contact Service if neces‐
sary.
b
Correct erroneous values.
DSP: NO RESP
Electronics communication problem (SR-unit).
b
Restart device; see above.
b
If not successful, contact Service.
EEPROM: CONTROL
Invalid microcontroller parameters (SR-unit).
b
Restart device; see above.
b
If not successful, contact Service.
EEPROM: LASER
Invalid laser parameters.
b
Restart device; see above.
b
If not successful, contact Service.
FIT: DIV BY 0
Error during signal evaluation:
•
Incorrect parameter values.
•
Hardware defect.
b
Check
measuring distance
,
substitute for tem‐
perature
and
pressure
parameters, correct when
necessary.
b
Restart device; see above.
b
If not successful, contact Service.
FIT: NO CONV
Error during signal evaluation:
•
Incorrect parameter values.
•
Hardware defect.
b
Check
measuring distance
,
substitute for tem‐
perature
and
pressure
parameters.
b
Increase
Average
parameter
a
Attention, these parameters may only be mod‐
ified by trained personnel.
b
Restart device; see above.
b
If not successful, contact Service.
FIT: S MATRIX
Error during signal evaluation:
•
Incorrect parameter values.
•
Hardware defect.
b
Check
measuring distance
,
substitute for tem‐
perature
and
pressure
parameters, correct when
necessary.
b
Restart device; see above.
b
If not successful, contact Service.
Incompatible device
Incompatible software (SR-unit).
b
Check software version.
b
Contact Service.
INIT: NO LINE
Incompatible software (SR-unit).
b
Restart device; see above.
b
If not successful, contact Service.
LD: PELT ERROR
LD…laser diode
Temperature measurement on Peltier element defec‐
tive:
b
Possible hardware defect.
b
Restart device; see above.
b
If not successful, contact Service.
LD: TEMP ERROR
Laser wavelength adjustment outside allowed range.
b
Absorption line “lost”.
b
Laser diode defective.
b
Restart device; see above.
b
If not successful, contact Service.
MEAS: M PLAUS
Measuring results from measuring channel not plausi‐
ble:
b
Incorrect parameter values.
b
Strong signal interference.
b
Hardware defect.
b
Check
measuring distance
,
substitute for tem‐
perature
and
pressure
parameters.
b
Increase
Average
parameter.
a
Attention, these parameters may only be mod‐
ified by trained personnel.
b
Restart device; see above.
b
If not successful, contact Service.
MEAS: R PLAUS
Measuring results from measuring channel not plausi‐
ble:
b
Incorrect parameter values.
b
Strong signal interference.
b
Hardware defect.
b
Restart device; see above.
b
If not successful, contact Service.
MEAS: REF CONC (only for
device with closed cell)
Gas concentration in reference cell too low.
b
X
Observe safety information for Ex-free zone!
b
Exchange reference cell.
b
Contact Service.
Sensor communication
SR-unit not connected correctly.
b
Check CAN connection and repair if necessary.
SIG: DARK VALUE
Dark values of receiver element too high:
b
Possible hardware defect.
b
Contact Service.
SIG: K HIGH
Monitoring channel signal too high.
b
Contact Service.
SIG: K LOW
Monitoring channel signal too low.
b
Contact Service.
SIG: M HIGH
Measuring channel signal too high.
b
Contact Service.
TROUBLESHOOTING
10
8019532/100Q/V1-1/2018-08 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM700 Ex
83
Subject to change without notice