74
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Terminal Modules (Accessory)
7.2.4
Installation
1
Secure the terminal module to a mounting rail.
2
Connect the 37-pole plug connector of the terminal module with the “Analog and digital
I/O” plug connector of the operating unit. To this purpose use the special connection
cable (order no.
):
Connect the “Connecting module” plug connector to the terminal module.
Connect the “Remote unit” plug connector to the operating unit.
3
Connect a 24 VDC power supply to terminal strip X2 (
).
4
Configure the terminal module as required (
).
5
Establish the signal connections accordingly (
7.2.5
Function of the relay switching contacts
Switching logic
Load capacity
Subject the switching contacts of terminal module 2 to a load of 24 V/500 mA max.
Only connect inductive loads if suppressor diodes (free-wheeling diodes) are provided.
7.2.6
Configuration options (jumper)
Connection of the analog outputs
In the standard configuration the analog outputs on terminal strip X20 are available; can,
however, also be connected to X10. The jumpers J4 and J5 serve for configuration
purposes:
The connector pin assignment of the connection cable are not identical.
Do not confuse the ends of the connection cable.
Logical function condition
Condition of the relay switching contact
Enabled
Closed
Disabled
Open
Power supply failure/switched off
Open
When the mains voltage is switched off, the alarm signals of the operating unit
are in the “Alarm” status.
NOTICE:
Inductive loads (e.g. relays, solenoid valves) can severely damage the relay
switching contacts if the induced voltages are not discharged through
suppressor diodes.
With inductive loads:
Check whether suppressor diodes are installed.
If this does not apply:
Install separate suppressor diodes.
Table 6
Jumper configuration for the analog outputs of terminal module 2
Use the analog output
on X20
on X10
1 Analog output 1 “Measured value”
J4 bridged
J4 open
2 Analog output 2 “Reference value”
J5 bridged
J5 open
The analog outputs are potential-free towards the other electronic
components, but not towards one another (joint minus pole).
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...