Calibration
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
151
12.5
Manual calibration
Function
Calibration can be started manually with the “Calibration” menu item, and the automatic
calibration procedure can be observed. The calculated drift values (“deviations”) can be
seen at the end of the calibration operation and you can decide whether these calibration
results should be accepted. If you refuse the results, the previous calibration remains
effective.
Procedure
1
Check whether the conditions for calibration are met:
– The EuroFID is ready for operation.
– The calibration gas is supplied.
– The setpoint value is set correctly (
).
2
Open the menu: Main menu
Calibration.
To start calibration:
3
Press .
➝
The “Calibration” status output is activated.
➝
The calibration operation runs automatically.
If the “Calibration block” control input is (
), no calibration
operations can be performed.
C a l i b r a t i o n
Z e r o G a s :
0 . 0 0 p p m
0 . 0 0 m g C
C a l . G a s :
7 . 9 1 p p m
1 2 . 7 2 m g C
r u n Y e s
C h a n g e o r
6 1 . 6 p p m
R a n g e . . . 1 0 0
Menu title
Zero gas setpoint value
1
Zero gas setpoint value
2
Span gas setpoint value
3
Span gas setpoint value
Selection box
Current measured value
Current measuring range
1
Not variable
2
In the selected physical unit
3
Setting
To terminate the function:
3
Press
or
.
➝
The selection box is highlighted.
4
Press .
➝
The selection box switches to
N O
.
5
Press
or
.
➝
E n d
is highlighted.
6
Press .
E n d
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...