The last valid distance value is frozen and output again (held). This process repeats
until it is once again possible to run a measurement. This behavior should only condi‐
tionally be used for safety-related purposes.
7.3.4.2
Laser
It is possible to switch off the sender laser temporarily or permanently. No measure‐
ment is possible when the laser is switched off.
Permanently switch the laser off
The laser can be switched off permanently via SOPAS ET. It remains switched off until it
is switched back on by another event. If the power supply of the device is interrupted in
the meantime, the laser remains switched off afterwards.
7.3.4.3
Multifunctional input
The laser can be switched off temporarily via the multifunctional input (MF) using the
Laser on/off
function. The laser switches off as soon as the switching condition has been
met and remains switched off until it is no longer met.
If the MF input is not used, the
Inactive
setting should be selected, or the input con‐
nected to GND or 24 V accordingly.