Electrical installation
6.1
Wiring instructions
NOTE
Pre-assembled cables can be found on the product page. which can be accessed via
the
SICK Product ID
pid.sick.com/
{P/N}
{P/N}
corresponds to the part number of the product (see type label).
NOTICE
Faults during operation and defects in the device or the system
Incorrect wiring may result in operational faults and defects.
■
Follow the wiring notes precisely.
The electrical connection of the device is configured as an M12 round connector.
The enclosure rating stated in the technical data is achieved only with a screwed plug
connector or protective cap.
All circuits connected to the device must be configured as SELV or PELV circuits. SELV =
safety extra-low voltage, PELV = protective extra-low voltage.
Protect the device with an external 2 A slow-blow fuse at the beginning of the supply
cable.
6.2
Connecting the device electrically
1.
Ensure the voltage supply is not connected.
2.
Connect the device according to the connection diagram,
3.
Switch on the supply voltage.
6.3
Connection diagram
Power
Table 1: Pin assignment Power connection
Male/female
connector
Contact
Short form
Signal description
M12 male con‐
nector, 5-pin A-
coded
1
4
3
5
2
1
L+
Supply voltage
2
Qa/Q2
Analog output / digital output 2
3
M
Supply voltage: 0 V
4
Q1
Digital output 1 / IO-Link
5
MF
Multifunctional input MF