OPERATING INSTRUCTIONS | DOSIC
®
8020955 /11HD/ 2018-12-12 | SICK AG
Subject to change without notice
40
9
OVERVIEW OF PARAMETERS
9
Overview of parameters
Parameter
Description
Q1/Q2-Flw
Q1 or Q2 set as flow digital output (Switch/Frequency/Puls)
Q1/Q2-Tmp
Q1 or Q2 set as temperature digital output
Q1/Q2-Sta
Q1 or Q2 set as status digital output
Q1/Q2-In
Q1 or Q2 set as digital input
Q1/Q2Mode
Selection of mode for Q1 and Q2. Values: Output (digital output)
or input (digital input)
Q1/Q2Act
Selection of action for Q1 and Q2. Values: ResetV (volume count
-
er set to 0) or Cutoff (deactivation of cut-off function)
Q1/Q2Proc
Selection of function for Q1 and Q2 as outputs. Values: Flow,
Temperature, or Status
Q1/Q2Out
Selection of set values for Q1 and Q2. Values: Hysteresis or
Window
Q1/Q2SP
Q1 or Q2 switching point value (if Q1 or Q2 is set as Flow or Tem-
perature and Hysteresis and Q2 is set as switch output mode). If
Q2 is set as a frequency output, the maximum flow is set under
maximum frequency.
Q1/Q2RP
Q1 or Q2 reset point value (if Q1 or Q2 is set as Flow or Tem-
perature and Hysteresis and Q2 is set as switch output mode). If
Q2 is set as a frequency output, the minimum flow is set under
minimum frequency.
Q1/Q2FH
Q1 or Q2 upper threshold (if Q1 or Q2 is set as Flow or Tempera-
ture and Window and Q2 is set as switch output mode).
Q1/Q2FL
Q1 or Q2 lower threshold (if Q1 or Q2 is set as Flow or Tempera-
ture and Window and Q2 is set as switch output mode).
Q1/Q2Pol
Switching output Q1 or Q2. Values: no or nc (normally open or
normally closed)
Q1/Q2Sim
Simulation mode Q1 or Q2. Values: SimOff (no simulation), Inactv
(0 or 24 V, QxPol-dependent), Active (0 or 24 V, QxPol-dependent)
Q1/Q2Type
Switching output mode selection Q1 or Q2. Values: PNP, DRV, OC,
NPN (PNP, push-pull, open collector, NPN)
Q1/Q2Stat
Sensor output. Values: Fail (system error), Empty (channel emp
-
ty), Steril (sterilization detection) Negatv (negative flow)
Q2Func
Q2 digital output type selection. Values: Pulse, Freq or Switch
PlsVal
Volume that generates a pulse when Q2 is set as pulse output
PlsWid
Pulse duration if Q2 is set as frequency output
FrqMax
Maximum frequency if Q2 is set as frequency output
FrqMin
Minimum frequency if Q2 is set as frequency output
Qa/Qb-Off
Qa or Qb is deactivated
Qa/Qb-Flw
Qa or Qb set as flow analog output
Qa/Qb-Tmp
Qa or Qb set as temperature analog output
Qa/QbType
Select if the analog output is active (4 mA ... 20 mA) or inactive
(Off)