FLOWSIC100 Flare · Addendum to Operating Instructions · 8025365/V1-0/2020-05 · © SICK Engineering GmbH
Installation
Fig. 20
Marking the nozzle positions on the pipeline for cross-duct versions
1-path measurement
2-path measurement
6) Roll the strip back around the pipeline and mark the nozzle positions with crossing and marking points by using a metal center punch.
7) Take the strip off again and join the additional markings with a line.
Crossing point (marking for
the center of the nozzle)
Marking point (aid for
aligning the nozzle)
Path 1
Path 2
Path 1
Path 2
Crossing point (marking for
the center of the nozzle)
Marking point (aid for
aligning the nozzle)
Marking lines
Marking lines