![SICK 8013344 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/sick/8013344/8013344_addendum-to-operating-instructions_1264413025.webp)
Product description
FLOWSIC100 Flare · Addendum to Operating Instructions · 8025365/V1-0/2020-05 · © SICK Engineering GmbH
25
3.3
Material for parts with process gas contact
Table 3
Overview of parts with process gas contact
Fig. 9
Parts with gas contact
Material
Component
Type FLSE100-XT
F1F-S F1F-M F1F-H F1F-P
Stainless steel
1.4571
Duct probe (9), transducer (7), connecting flange (6)
x
Connection for optional venting (10), retraction nozzle (5)
x
Duct probe (9), connecting flange (6), sensor contour, probe version
F1F-P (12)
x
Adjusting ring (14), pipe screw fitting (1), sealing disc (3)
x
x
x
x
Titanium
Duct probe (9), transducer (7)
x
Transducer (7), transducer holder within sensor contour (13)
x
PTFE
Centering (11)
x
x
x
x
Sensor contour (8)
x
Thrust ring (15)
x
PTFE/graphite
Sealing profile (4)
x
x
x
x
Stainless steel
1.4568
Spring (2)
x
x
x
x
11
3
2
10
4
14
5
6
9
15
1
13
7
11
8
1
4
10
9
7
2
6
5
3
1 Pipe screw fitting
9 Duct probe
2 Spring
10 Connection for optional venting
3 Sealing disc
11 Centering
4 Sealing profile
12 Sensor contour, probe version F1F-P
5 Retraction nozzle
13 Transducer holder within sensor contour
6 Connecting flange
14 Adjusting ring
7 Transducer
15 Thrust ring
8 Sensor contour
Cross-duct versions
Probe version
12