The output voltage is applied to the device under test 

through the source impedance. The voltage across 

the device is always less than or equal to the
output voltage. Source impedance values are 

62.5

Ω

 (R1), 1k

Ω

 (R2) and 16 k

Ω

 (R3). 

The source impedance is selected as a function of the measurement 

range. For most devices, including resistors, most capacitors and many 

inductors, the 1.0 Vrms setting provide best result. For some inductors 

and active devices, the 0.25 or 0.5 Vrms setting should be used. 
Certain devices require a speci

fi

 c test voltage, such 

as Z5U ceramic capacitors (test voltage = 0.5 Vrms 

for 25V parts and 1.0V for < 16V parts).
Note: Use the largest voltage possible for the best SNR and accuracy.
The ST5 provides three measurement ranges (R1-R3). Each of 

the three ranges has the source impedance of approximately 

the mid-scale impedance. The table below speci

fi

 es the 

impedance ranges for each of the measurement ranges.

Range

Source R

DUT Impedance

1

62.5

Ω

< 400

Ω

2

1 k

Ω

400 

Ω

 <  < 4 k

Ω

3

16  k

Ω

> 4 k

Ω

Note: 

Measurement ranges determine an impedance range (not 

a value range), so the ranges for inductance and capacitance 

depend upon the test frequency. In addition, the impedance 

for capacitors is inversely proportional to its capacitance, so 

larger capacitors are measured in lower impedance ranges.
During normal operations the unit automatically changes to the 

optimal range for the DUT. There is a built-in hysteresis allowing to 

avoid repeated range changes when a component is near a range 

boundary. The auto-ranging function can be disabled by user.
Range holding is helpful if a number of parts with similar values are 

being measured or a speci

fi

 c measurement condition required.

5

Содержание st5

Страница 1: ...3...

Страница 2: ......

Страница 3: ...ntrols 6 PowerOn 7 MenuStructuresAndFunctions 9 MeasurementFeatures 20 MeasuringResistance 20 MeasuringCapacitance 20 MeasuringInductance 21 Maintenance 22 Labelling VerificationRequirements 23 Append...

Страница 4: ...ITY OF THE BATTERIES OR PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO FOLLOW INSTRUCTIONS THIS WARRANTY DOES NOT COVER LCD DAMAGE PHYSICAL DAMAGE TO THE JOG DIAL BUTTON SLIDE SWITCH AND RESET SWITCH...

Страница 5: ...sing Donotusethedeviceifitappearstobedamaged Do not use the device if it operates abnormally Do not attempt to measure any components in circuit when your circuit is alive or active Toavoidpossibledam...

Страница 6: ...ice STiscapableofmeasuringresistance capacitance orinductanceoverarangeofmorethan8ordersof magnitude Thedevicehasabasicaccuracybetterthan 0 2 resistance andoperatesatfourtestfrequencies SmartTweezersi...

Страница 7: ...ntconversionresistorRi The outputofAIprovidesasignalproportionaltothecurrent I Ri VoltageacrosstheDUTismeasuredbyaseparatesignalpath amplifierAU thusprovidingapseudo4 wireKelvinconnection Voltageandcu...

Страница 8: ...asurementranges R1 R3 Eachof thethreerangeshasthesourceimpedanceofapproximately themid scaleimpedance Thetablebelowspecifiesthe impedancerangesforeachofthemeasurementranges Range SourceR DUTImpedance...

Страница 9: ...ontrolsallowchangingtestparametersormodes withoutenteringthegeneralmenubymovingtheNavigation ControllerUP DOWN LEFTandRIGHTasshownbelow Thefollowingtablesummarizesfunctionsof theNavigationControllermo...

Страница 10: ...ncontrolleris operatedforapproximately30seconds defaultvalue Thepowerofftimeoutvaluecanbesetbychanging theTIMEOUTsettingintheSYSTEMmenu Thedefaultpower offtimeoutis30secondsina measurementmodeand30sec...

Страница 11: ...odeindicateAuto Resistance Inductance Capacitance ImpedanceandESR measurementandDiodeTestmoderespectively DISPLAYEDPARAMETERS Themeasurementmodesetting R L R C R C D L Q Z ESRandAUTO determinesthemeas...

Страница 12: ...eal and thusalowD Z MODE TheImpedanceofthecomponentisshown onPrimaryDisplay Unitsarem k orM ESRMODE Theequivalentseriesresistanceofthecapacitoris shownonthePrimaryDisplay ESRunitsarem k orM AUTOMODE S...

Страница 13: ...erUPorDOWNtomovecursor tothedesiredmenuitemandPRESSittoselecttheitem TheCurrentSettingcursorindicatesthecurrentsetting MAINMENU MainmenuisusedtoaccessSystemmenu Measurementmenuorto restoremeasurementp...

Страница 14: ...esoundforallfunctionsexceptforthe NavigationControlleroperation SelectR TONEtoenableaspecialmodewhenbeepfrequencyvaries dependingonthemeasuredresistancevalueintheResistanceMode seetheMeasurementMenuse...

Страница 15: ...adjustdisplaycontrast MoveNavigation ControllerUPorDOWNtochangecontrast PRESS toexitmenuattheadjustedcontrastlevel TIMEOUT SelectTIMEOUTtoadjustthetimeoutbeforetheunitgoesto sleepmode MoveNavigationCo...

Страница 16: ...SUREMENTMENU Measurementmodesandsettings MODEMENU TheModemenuisusedtosetthemeasurementmode SelectRES IND CAP IMPorESRmenuitemstomeasuredesirablecomponentor parameterasResistance Inductance Capacitance...

Страница 17: ...tion INDUCTANCEMODE EnablesCapacitancemeasurementmode SeesectionMEASUREMENTFEATURESformoreinformation CAPACITANCEMODE EnablesCapacitancemeasurementmode SeesectionMEASUREMENTFEATURESformoreinformation...

Страница 18: ...stfrequency RDQMENU Usethismenutosetsecondarydisplayparameter Thefollowingcombinationsareallowed C R capacitance resistance C D capacitance dissipation factor L R inductance resistances L Q inductance...

Страница 19: ...esistive componentinseriesorinparallelwithareactivecomponent Dependinguponthecharacteristicsofthecomponenttheseriesor parallelmodelwillbemoreaccurate Inmostcases partsarebest approximatedbytheseriesmo...

Страница 20: ...SmartTweezerswilldisplaydifferenceinpercentfrom thereferencevalueandthebeeperwillbeep 1 time when the component is within the setting tolerance 3 times whenever the component under test exceeds the se...

Страница 21: ...t value by measuring a component or by nulling test leads see examples below Enter the NULL menu and select SET Example1 Nullingtestleadsforsmallresistancemeasurement Select manual R measurement mode...

Страница 22: ...D mode see HOLD menu Enter the NULL menu and select ZERO ToresettheNULLmodecompletelyselectAUTOSETfrom themainmenuorDEFAULTfromtheSETTINGSmenu HOLDMENU Allowstoholdlastreadingondisplay PERIODMENU Peri...

Страница 23: ...n25m andmay increaseifthegoldonthetweezertipswearsout Theoffset valueshouldbeusedincalculationoftheactualresistance MEASURINGCAPACITANCE Test frequency 0 1kHz 1 kHz 10kHz Test signal amplitude 0 25 0...

Страница 24: ...s Componentsize Offset pF 1206 0 58 0805 0 6 0603 0 65 0402 0 7 MEASURINGINDUCTANCE Test frequency 0 1kHz 1 kHz 10kHz Test signal amplitude 0 25 0 5 1 0 Vrms Sine wave Source impedance 62 5 1k 16k ohm...

Страница 25: ...istors capacitors andinductors haveparasiticcomponents Thus simplecomponents shouldbemodeledascompleximpedances Test frequency 0 1kHz 1 kHz 10kHz Test signal amplitude 0 25 0 5 1 0 Vrms Sine wave Sour...

Страница 26: ...e performedinordertotroubleshoottheproblem 1 Check battery voltage and recharge if necessary 2 Review this manual for possible mistakes in the operating procedure 3 Reset device by reconnecting batter...

Страница 27: ...ve Source impedance 62 5 1k 16k 1 TYPICALOFFSET Resistance 25 m Capacitance 0 65 pF Inductance 0 1 uH Offsetvalueshouldbesubtractedfrommeasurementresultfor smallvaluecomponents R 10 C 100pF L 10 H MEA...

Страница 28: ...Phase angle F 0 1 deg indicationoftheparameternotimplementedinsome versions Auto mode Read out Dominant parameter Equivalent circuit diagram Serial Parallel for C R Serial for L R Manual Mode Read ou...

Страница 29: ...No change APPENDIXC ACCURACYSPECIFICATION RESISTANCE IMPEDANCE Range Resolution 100Hz 1kHz 10kHz 1R 0 001R 0 7 50 0 7 50 0 7 50 10R 0 01R 0 7 8 0 7 8 0 7 8 100R 0 01R 0 2 3 0 2 3 0 2 3 1000 0 1R 0 2...

Страница 30: ...5 0 2 3 0 5 3 1000pF1 0 1pF NA NA 0 5 5 0 5 3 100pF1 0 01pF NA NA 0 5 10 0 8 20 10pF1 0 001pF NA NA NA 1 0 50 Accuracyfortherangesof10pF 1000pFiSspecifiedaftersubtractofthestray capacitancesfortestle...

Страница 31: ......

Страница 32: ...2...

Отзывы: