21
DUT
Optimal test frequency
<10000pF
10 kHz
10001pF- 1
μ
F
1 kHz
There is some small capacitance offset due to capacitance of the tips. The
offset depends on the distance between the tips (i.e. measured component
size). The offset value should be used in calculation of the actual capacitance.
Table below shows typical offset values for different component sizes:
Component size
Offset, pF
1206
0.58
0805
0.6
0603
0.65
0402
0.7
MEASURING INDUCTANCE
Test frequency
0.1kHz/1 kHz/10kHz
Test signal amplitude
0.25/0.5/1.0 Vrms Sine wave
Source impedance
62.5/1k/16k ohm
Test period
1 Sec (default)
Equivalent circuit diagram
Serial
In AUTO mode ST automatically selects the best test frequency and is
capable of measuring inductance from 1
μ
H to 1kH. To measure inductance
lower than 5
μ
H or more than 500mH select test frequency manually:
DUT
Optimal test frequency
<100
μ
H
10 kHz
100
μ
H -100 mH
1 kHz
> 100 mH
100 Hz