19
© Copyright Siborg Systems Inc.
Press joystick to displays Serial Number of the device and the
Firmware version.
The following settings are stored in the device EEPROM when
Default is selected in the System Menu and the joystick is
pressed:
•
Primary Parameter
Auto
•
Secondary Parameter
Auto
•
Frequency
Auto
•
Signal Level
1.0 Vrms
•
Measurement Period
0.5 s
•
Equivalent Circuit
Auto
•
Sound
On
•
Data Hold
Off
•
Shutdown Time
120 s
•
Wakeup Mode
R-L-C-D
•
RDC+LED Mode
Auto
•
Frequency Mode
Frequency
•
Frequency Measurement Time
1 s
•
System Sound
Medium
•
All offsets are reset to zero
8.1 If the device does not turn on
•
Push the joystick and hold it for 1 beep, then release it.
•
Charge the battery.
•
Contact the manufacturer for repair.
8.2 If there are no initial readings on the screen
•
From the system menu, select Default.
•
Clean the contacts and make open/short calibration with open
and closed probes.
•
Review this manual to find possible errors in operating the
device.
Customers can contact customer support by phone +1-519-888-
9906 or by e-mail [email protected].
When contacting technical support, please provide the following
information:
•
Model number
•
Software version number
•
Serial number of the device
•
Purchase Receipt