Chapter 7: Spirometry Technique
DATOSPIR MICRO User’s Manual
201
511-A00-MU2 • REV. 2.06
shown to be similar or to have equivalent results. The volume
extrapolated using this method must be below 5% of the vital
capacity or 150 ml, without exceeding either of the two criteria.”
“To ensure good spirometry, the technician will pay particular
attention to ensure that the patient has made the utmost effort,
that the start has been good and that no coughing or Valsava’s
manoeuvre due to glottis closure has occurred. Special attention
must be paid to preventing expiration from ending too soon,
which would be detected in the end of the curve, which would
be too perpendicular to the horizontal base line. Sometimes
the patient inadvertently partially obstructs the mouthpiece
with his tongue or false teeth. It is essential that the best two
expirations from the best three acceptable curves do not vary by
more than 200 ml of FVC or FEV1. The best effort cannot only
be determined by merely inspecting the spirometric curve but
also by checking the measurements to determine the maximum
values. The independent selection of FVC and FEV1 sometimes
causes greater variability, as factors such as learning, tiredness or
bronchial spasms caused by expiration come into play. The best
FEV1 does not have to be rejected when the manoeuvre from
which it originates has ended prematurely. However, FEF25-75%
is influenced by the vital capacity of the curve chosen. Falsely
high values may be recorded if a manoeuvre with a shortened
or lower vital capacity than the real vital capacity of the patient
has been chosen. It seems that the most practical criterion is
to choose the manoeuvre containing the FVC and FEV1 with the
highest sum among the three chosen for calculation.”
7.2 CALIBRATIONS
“As well as the calibration procedures included by the manufacturer
in the appliance to quickly check the working order of the basic
pneumotachometer circuits and mechanisms, it must be possible
to check the appliance through external signals. These signals
must be as similar as possible in terms of flows, volumes and
times to the biological signal for which the instruments were
designed, i.e. forced expiration. This is not always possible,
Содержание DATOSPIR MICRO A
Страница 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Страница 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Страница 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Страница 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Страница 31: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 31 511 A00 MU2 REV 2 06 2 MODEL A OPERATION...
Страница 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Страница 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Страница 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Страница 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Страница 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Страница 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Страница 171: ...Chapter 4 Communications system DATOSPIR MICRO User s Manual 171 511 A00 MU2 REV 2 06 4 COMMUNICATIONS SYSTEM...
Страница 177: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 177 511 A00 MU2 REV 2 06 5 TECHNICAL SPECIFICATIONS...
Страница 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Страница 193: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 193 511 A00 MU2 REV 2 06 6 OPERATING PRINCIPLES...
Страница 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Страница 199: ...Chapter 7 Spirometry Technique DATOSPIR MICRO User s Manual 199 511 A00 MU2 REV 2 06 7 SPIROMETRY TECHNIQUE...
Страница 212: ...DATOSPIR MICRO User s Manual 212 Chapter 8 Upkeep Preventative and Corrective Maintenance 511 A00 MU2 REV 2 06...
Страница 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...