background image

Well sounders «DU-1» 

 

outer  surfaces  with  a  brush.  Do  not  leave  the  valve  in  the  washing  liquid 
for a long time. After washing you should wipe the valve dry with a clean 
rag. Before installing it back in the well sounder, apply a small amount  of 
lubricant,  according  to  standard  CIATIM  GOST  6267-80,  on  the  rubber 
sealing ring surface (5). 

After  mounting  the  valve  on  the  well  sounder,  install  the  latter  on 

the  Inspection  bench  for  well  sounders  SKU-1  IZM  4.137.003  (SIAM 
Company)  and  purge  the  valve  at  pressure  of  5…8  atm,  using    clean  air. 
Test  for  pressure  integrity  with  a  soap  solution.  A  slight  “seepage”  is 
allowed  with  small  bubbles  in  the  discharge  hole  area.  In  case  of  a  high 
"seepage", repeat the valve washing process. 

If high “seepage” continues, then replace the cone (7) of ball seats. 

In  order  to  do  that,  press  the  retaining  ring  (3)  and  pull  it  out  of  valve 
housing. Take out the guide bush (1), spring (4), tappet (2) in combination 
with  cone  (7)  and  nut  (6)  from  valve  housing  (8).  Then,  replace  the  cone 
(7) by unscrewing the nut (6). When necessary, clean the contact chamfer 
of the valve ball-and-seat in the housing (8) with a rag. 

Assemble it in the reverse sequence.

 

11.2.3. Cleaning of Accumulator Charging Socket 

Cleaning  is  done  with  a  pure  ethyl  alcohol  or  alcohol-gasoline 

blend  (1  part  of  ethyl  alcohol,  1  part  of  rubber  solvent  or  oil  solvent 
gasoline, the other kind of gasoline is not allowed), using a brush. Alcohol 
consumption  –  30g  per  month  of  operation.  Apply  a  small  amount  of 
CIATIM GOST 6267-80 lubricant on the connector surface after cleanup. 

11.2.4. Performance Monitoring 

Well  sounder  performance  control  is  done  at  the  SIAM  service 

center or by SIAM certified specialists, according to design documentation 
under the testing program. 

11.2.5. Pressure Testing 

Well sounder pressure testing is done at the SIAM service center or 

by  SIAM  certified  specialists  on  the  Hydraulic  testing  bench  SGI-1  IZM 

Содержание DU-1

Страница 1: ...entific Industrial Introduction Company SIAM LLC WELL SOUNDERS DU 1 OPERATING MANUAL Explosion proof design type of explosion protection intrinsically safe circuit 1Ex ib IIB T3 Gb explosion proof mar...

Страница 2: ...Well sounders DU 1...

Страница 3: ...OFNESS OF THE PRODUCT 13 7 EQUIPMENT INSTALLATION AND DEINSTALLATION ORDER 16 7 1 Preparation of Test Objects for Echometry 16 7 2 Well sounder Installation 16 7 3 Well sounder Removal 17 8 BATTERY CH...

Страница 4: ...32 11 2 6 Replacement of the Valve Sealing Ring 33 12 DEVICE STORAGE AND TRANSPORTATION 34 APPENDIX 1 Control of Well Fluid Level in Complicated Conditions 35 APPENDIX 2 Setting Sound Speed Work Shee...

Страница 5: ...range from 40 C to 50 C The device design is comply with the assigned explosion proof mark with the requirements of GOST IEC 60079 14 2013 and other regulatory documents governing the use of electric...

Страница 6: ...31610 2011 and have 1Ex ib IIB T3 Gb explosion proof mark DU 1 well sounder is designed for indoor and outdoor installation in hazardous areas where atmospheric explosives of IIA IIB categories and T1...

Страница 7: ...r consists of 1 the case 2 thread connecting coupling 3 two shackles 4 the rotary retrievable outlet valve 5 the ON OFF button 6 the Power LED 7 the Charge LED 8 the plug for charging the in built acc...

Страница 8: ...e maintains its operability at ambient temperatures from minus 40 C to 50 C 2 3 The device operates autonomously and is powered by a special internal battery type EEMB LP103450LC PCM LD with 3 6 V or...

Страница 9: ...ing to the set shown in Table 2 Table 2 Title Q ty 1 Well sounder DU 1 1 2 Certificate 1 3 Operating manual 1 4 DB Siam 2 5 software user s guide 1 5 CD with the software 1 6 Package 1 7 Conformity ce...

Страница 10: ...rmed After the acoustic impulse is generated the rubber sphere should be kept in the position which it took immediately after the blow until the value of the measured level pops up on the screen The b...

Страница 11: ...y approved by the RTSA 5 2 Requirements to Personnel Well tests are to be carried out by individuals above 18 who have secondary education and who are medically cleared for work Personnel are to take...

Страница 12: ...tapers chokes hoppers and similar units The elbow of the offset should have pipe tapered thread 60 State Standard 633 80 and be located at the height 0 2 1 8 meters above the ground If the height exc...

Страница 13: ...ces The minimum width of conductors on the PCB is 0 2 mm copper thickness is not less than 18 m Thus the electrical circuit coming out of the battery compartment of the device is intrinsically safe Th...

Страница 14: ...hall be replaced by new ones supplied by the manufacturer 6 The device s operability is maintained at ambient temperature from minus 40 to plus 50 The maximum temperature of overheating of components...

Страница 15: ...osion proofness during operation Check that the cover of the battery compartment and the compartment itself including the threaded connection are free from mechanical damage and corrosion In case of c...

Страница 16: ...nnecting coupling of the sensor If the thread segment is worn out or significantly corroded or damaged well sounder installation is prohibited Open the valve on the offset for a short period of time 1...

Страница 17: ...on the terminal indicator In order to avoid leaks in thread connections use 402 dope 38 101 330 73 or 0 1 20 sealing band 6 05 1388 86 7 3 Well sounder Removal Switch off the device Close the valve of...

Страница 18: ...via an appropriate cable to a network or car adapter or USB port of your computer The charging time of a fully discharged battery is approximately 7 hours The voltage of a fully charged battery is 4 1...

Страница 19: ...mputer you must connect the sensor using Windows Add Bluetooth device The procedure for connecting may vary depending on the OS version The sensor must be turned on and ready for connection The sign o...

Страница 20: ...ck the right mouse button in the appeared context menu select item Open parameters see Figure 1 in the appeared Bluetooth Options window on the COM Ports tab see which outgoing port is assigned to the...

Страница 21: ...ers DU 1 Working with devices in DB SIAM To work with devices in the DB SIAM software start the program switch to the Settings tab press the Sensors button see Fig 3 the Sensors dialog box will appear...

Страница 22: ...blems a corresponding warning message appears This happens if the COM port is incorrectly specified or in case of the sensor is switched off If the error occurs because the sensor has been switched of...

Страница 23: ...Well sounders DU 1 If the connection is successful the sensor control panel appears For each sensor type the control panel will have a different look...

Страница 24: ...e sensor and allows the operator to save the study into the database at the moment when he is sure that he has correctly specified the well details and other study parameters If the operator has not s...

Страница 25: ...ed to specify what sound speed to use for level detection You can use one of the tables of correction tables of dependence of sound speed in annular space on pressure in annular space or set the value...

Страница 26: ...dy to start state and press the Start button 6 The Waiting for burst state would succeed the state Noise Measurement 7 Run the burst 8 The device will switch to the state of recording the echogram the...

Страница 27: ...e 10 2 Multi purpose Valve The multi purpose valve is supplied separately and is to be used in SUDOS well sounders and SIAM manufactured well sounders for generating an acoustic impulse by injecting g...

Страница 28: ...that allows to connect the conversion kit to the well sounder receiver The cylindrical receiver from GBO 2 kit is screwed on the outlet of the automatic valve unit or on the specialized manual valve o...

Страница 29: ...12 months Cleaning of the tapered connecting thread the acoustic sensor the well sounder and the case Maintenance of the exhaust valve Cleaning of accumulator charging socket Performance monitoring 1...

Страница 30: ...or When cleaning a very dirty thread application of metal tools an awl a screwdriver etc is allowed but do not apply an excessive force not to damage the thread It is strictly forbidden to use a metal...

Страница 31: ...nents are given in the Figure 1 guide bush 2 tappet 3 retaining ring 4 spring 5 sealing ring 6 nut 7 cone 8 housing Washing is done with diesel fuel or kerosene using a brush and a rag Before washing...

Страница 32: ...f valve housing Take out the guide bush 1 spring 4 tappet 2 in combination with cone 7 and nut 6 from valve housing 8 Then replace the cone 7 by unscrewing the nut 6 When necessary clean the contact c...

Страница 33: ...with CIATIM GOST 6267 80 lubricant Then the valve should be put in place After a final assembly of the well sounder it is necessary to carry out the integrity and strength tests First install the wel...

Страница 34: ...less than 80 Product transportation is allowed in a standard bag by any type of transport at the temperature from 50 to 50 Severe shock and vibration are to be avoided during transportation When stor...

Страница 35: ...f the original signal and usually difficult to read In the annular space a level of interference is very high what is caused by an excessive vibration of the tubing hanger and casing ESP vibration val...

Страница 36: ...of high interference are acoustic noises on the part of flowline In this case it is recommended to shut a casing valve of the flowline tree during measurement if it is not in conflict with the safety...

Страница 37: ...in one field range The device will show the level based on the preset sound speed General use sheets are recorded in the read only memory of the device obtained from averaged data for respective regio...

Страница 38: ...344 6 1 361 10 9 378 0 8 328 3 0 345 6 3 362 11 2 379 1 0 329 3 1 346 6 6 363 11 5 380 1 1 330 3 3 347 6 9 364 11 9 381 1 2 331 3 5 348 7 1 365 12 3 382 1 3 332 3 7 349 7 5 366 12 8 383 1 5 333 3 9 3...

Страница 39: ...speed m s 0 7 310 2 8 327 6 3 340 13 5 351 0 8 311 3 0 329 6 6 341 14 0 352 1 0 313 3 1 329 6 9 341 15 0 353 1 1 314 3 3 330 7 1 342 17 0 354 1 2 315 3 5 331 7 5 343 20 0 356 1 3 316 3 7 332 7 7 343...

Страница 40: ...E mail siamoil siamoil ru Web address http www siamoil ru For repair and maintenance of products manufactured by TSIIC SIAM LLC contact the service centers of the company 1 Service center in Neftejug...

Страница 41: ...Well sounders DU 1 4 Service center in Tomsk 3 Belaya Str Tomsk 634003 Russian Federation tel 3822 90 00 08 20 02...

Страница 42: ...Well sounders DU 1 REVISION RECORD SHEET Revi sion Numbers of sheets pages Total sheets pages of document Ref of docu ment Incoming of supporting document Signa ture Date Revised Replaced New Annulled...

Отзывы: