background image

ver.210922E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

4.

 

Usable Conditions 

 

Usable solvents are listed below.   
Chloroform, Tetrahydrofran (THF), Toluene, Dichloroethane, Ethylacetate, N,N-Dimethylformamide (DMF) and 
Acetone 

 

 

Attention!

   

· Use the column within above stated flow rate, pressure, and temperature ranges. Using 

the column outside the given range may damage the column and lower its performance.   

 

·

 

Column pressure is influenced by the eluent composition, flow rate, and column  

temperature. When changing the eluent compositions, adjust the flow rate and column 
temperature so that the column pressure remains below the usable maximum pressure.   

 

· Shear degradation is more likely to occur in larger molecular weight compounds. When 

shear degradation occurs, the measured molecular weight may be lower than the actual 
value or it may influences the repeatability of the analysis. If shear degradation is 
suspected, use a lower flow rate.   

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Use HPLC grade organic solvent whenever possible. Solvents left in an opened bottle for 

a long time should not be used. The content may have been changed, absorbed moisture, 
or has been contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

   

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 

6. Sample Preparation 

(1) If possible, use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which 

has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes 
the sample. 

 

(2) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(3) Recommended injection volume is 500 - 2,000 

μ

L per column.   

 

(4) Viscosity of high molecular weight compound is largely influenced by its molecular weight and concentration. 

Samples with high viscosity cause peak broadening and elution delay, and this makes it difficult to obtain 
their accurate molecular weight distributions. In general, the larger the molecular weight of the compound, 
the higher its viscosity becomes. To suppress the influence from high viscosity, it is recommended to lower 
the sample concentration. Please use the below table as a reference when preparing samples for molecular 
weight distribution analyses. 

 

 

Molecular Weight Range 

Optimal Concentration (w/v) 

 5,000 

 1.0 % 

5,000 -   25,000 

 0.5 % 

 

(5) When sample contains water, methanol, or hexane, remove them prior to the injection.     

 

       

Note   

 

· Use of guard column is recommended to protect the analytical column. 

 
 
 

Product Name 

Flow Rate (mL/min) 

Maximum Pressure 

(MPa) 

(Per Column)

 

Temperature 

Range (°C) 

Recommended Maximum 

GPC FP-2002 

5 - 10 

20 

20 - 60 

GPC FP-G 8B

 

- - 

Содержание GPC FP-2002

Страница 1: ...f product name and serial number written on the column package column adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex GPC FP 2002 is a high performance fast preparative column for organic size exclusion chromatography It is suitable for fast preparative isolation of organic matter oligomer and polyme...

Страница 2: ...r has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is recommended 6 Sample Preparation 1 If possible use the eluent for analysis to dissolve or dilute samples If this is difficult use a solvent which has a composition that is as clos...

Страница 3: ...ween the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at less than 5 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the ...

Страница 4: ... and reagents being deposited inside the column These problems may be resolved by cleaning the column If a guard column is used with the analytical column first remove the guard column and check the performance of the analytical column alone If the problem is solved most likely the cause is from the guard column In this case clean the guard column If the problem is not solved by removing the guard...

Страница 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: