7.3 Solvent Exchange   

(1) When replacing the solvent, start the system at less than 0.5 mL/min.   

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3) When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 

use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

(4) When using a gradient method, changes in the eluent compositions may increase the column backpressure. 

Adjust the flow rate and column temperature so that the column backpressure remains below the usable 
maximum pressure.

 

 

7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column. When washing the column, let the washing solution flow 
from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate at less than 0.5 mL/min. Introduce 5 to 10 times column volume of the washing solvent. 

 

Method 1: Adsorption of ionic compounds   
Introduce an eluent with high salt concentration or 10-mM aqueous sodium hydroxide solution.   

 

Method 2: Adsorption of hydrophobic compounds   
Introduce a mixture of solvent containing 50-mM eluent (buffer or aqueous salt solution) and 50 % (v/v) 
acetonitrile or methanol.   

 

 

Attention!

   

· Keeping alkaline solution in the column for a long time will lead to the column deterioration. 

Please replace the washing solution with the eluent immediately after cleaning.

 

 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание Asahipak ES-502N 7C

Страница 1: ...umber written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak ES 502N 7C is a week anion exchange column filled with hydrophilic polymer base gel modified with diethylaminoethyl group It is an ideal column for analyzing various proteins peptides and components of nucle...

Страница 2: ... precipitation of salt When using highly corrosive salts such as sodium chloride wash out the salts at the end of analysis The metal parts of the devices and or the columns may rust Column pressure is influenced by the eluent composition flow rate and column temperature When changing the eluent compositions adjust the flow rate and column temperature so that the column pressure remains below the u...

Страница 3: ...stallation 1 Connect the column to LC system by following the flow direction arrow indicated on the column name tag 2 Make sure to insert the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide p...

Страница 4: ...tector Cleaning Method 1 Insoluble components that block the column inlet may be removed by reversing the flow direction i e introducing the eluent from the column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at less than 0 5 mL min Introduce 5 t...

Страница 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: