Shodex Asahipak ES-502N 7C Скачать руководство пользователя страница 2

ver.220216E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

4.

 

Usable Conditions 

 

Usable solvents are listed below. 
(1) The standard eluents are buffers and aqueous solutions of salt such as sodium chloride, potassium chloride, 

sodium sulfate, and potassium sulfate. Salts may also be added to the buffers. The recommended total salt 
concentration range is 20 - 600 mM. Select a buffer which has positive charge ion and has a large buffering 
capacity at the set pH. 

 

pH Range 

Buffer 

4.8 - 5.0 

N-Methylpiperazine·HCl 

5.0 - 6.0 

Piperazine·HCl 

5.5 - 7.3 

Bis-Tris

*1

·HCl 

6.4 - 7.3 

Bis-Tris Propane

*2

·HCl 

7.3 - 7.7 

Triethanolamine·HCl 

7.5 - 8.0 

Tris·HCl 

8.0 - 8.5 

N-Methyldiethanolamine·HCl 

8.4 - 8.8 

Diethanolamine·HCl 

8.5 - 9.0 

1,3-Diaminopropane·HCl 

9.0 - 9.5 

Ethanolamine·HCl 

 

*1

 2-bis(2-hydroxyethyl amino-2-(hyroxymethyl)-13-propanediol) 

*2

 1,3-bis(tris(hydroxymethyl)methylamino)propane 

 

(2) A mixture of buffer (or aqueous salt solution) and methanol or acetonitrile of any ratio can be used. Up to 

100 % methanol or acetonitrile can be used.   

 

 

Attention!

   

· Use the column within above stated flow rate, pressure, and temperature ranges. Using 

the column outside the given range may damage the column and lower its performance.   

 

· When using a mixture of buffer or salt solution and organic solvent, make sure there is no 

precipitation of salt. 

 

· When using highly corrosive salts such as sodium chloride, wash out the salts at the end 

of analysis. The metal parts of the devices and/or the columns may rust.

 

 

 

·

 

Column pressure is influenced by the eluent composition, flow rate, and column  

temperature. When changing the eluent compositions, adjust the flow rate and column 
temperature so that the column pressure remains below the usable maximum pressure.   

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

 

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 

Product Name 

Flow Rate (mL/min) 

Maximum 

Pressure (MPa) 

(Per Column) 

pH 

Range 

Temperature 

Range (°C) 

Recommended Maximum 

Asahipak ES-502N 7C 

1.0 

1.5 

1.2 

2 - 12 

10 - 50 

Содержание Asahipak ES-502N 7C

Страница 1: ...umber written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak ES 502N 7C is a week anion exchange column filled with hydrophilic polymer base gel modified with diethylaminoethyl group It is an ideal column for analyzing various proteins peptides and components of nucle...

Страница 2: ... precipitation of salt When using highly corrosive salts such as sodium chloride wash out the salts at the end of analysis The metal parts of the devices and or the columns may rust Column pressure is influenced by the eluent composition flow rate and column temperature When changing the eluent compositions adjust the flow rate and column temperature so that the column pressure remains below the u...

Страница 3: ...stallation 1 Connect the column to LC system by following the flow direction arrow indicated on the column name tag 2 Make sure to insert the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide p...

Страница 4: ...tector Cleaning Method 1 Insoluble components that block the column inlet may be removed by reversing the flow direction i e introducing the eluent from the column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at less than 0 5 mL min Introduce 5 t...

Страница 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...