background image

 

LBL 61876 rev. E, March 2018  

EN 

 

 

Step  

Picture or additional info if applicable 

 

Step 9 – Deliver Therapy 

While observing balloon positioning and lesion characteristics under fluoroscopy, press and hold the THERAPY button on the IVL Connector 

Cable. The IVL Generator will deliver lithotripsy pulses via the IVL Catheter balloon while the THERAPY button is depressed unless the IVL 

Generator determines that therapy is to be interrupted. As each therapy pulse is delivered, the THERAPY STATUS indicator will blink once, the 

PULSE COUNT display will decrement by one, and the Generator will sound one click. Confirm therapy delivery by continuously monitoring under 

fluoroscopy (see IVL Catheter IFU for additional information). To stop therapy, simply release the THERAPY button. 

 

NOTE: There is no need to make any adjustments for dosage levels or pulse rates.  All such settings are pre-programmed for given catheter 

types. 

 

 

 

Step 10 – Pause Period / Resume Therapy 

The IVL Generator is designed to force a brief pause in therapy at designated intervals.  If the user attempts to deliver a quantity of pulses 

without pausing, the IVL Generator will automatically interrupt therapy for a pause period.  The THERAPY STATUS indicator will be yellow during 

this period. To resume therapy, wait for the THERAPY STATUS indicator to become green again (two beeps will sound).  Simply release and 

press the THERAPY CONTROL again to resume therapy. Refer to the applicable IVL Catheter IFU for specifics on the maximum number of 

continuous pulses allowed and the duration of the pre-programmed pause period. Care must be taken not to exceed the recommended 

maximum number of pulses in the same treatment segment. 

 

 

 

Step 11 – IVL Catheter End of Life 

The IVL Generator is designed to sense the end of the useful life of the IVL Catheter.  Should this occur the PULSE COUNT will indicate “0” 

pulses remaining and the IVL Generator will interrupt therapy. The display will indicate a catheter error and a yellow light will appear around the 

POWER ON/OFF button. Replacing the IVL Catheter with a new one is required before the IVL Generator may be used again. Refer to the 

applicable IVL Catheter IFU for maximum pulse count per catheter (useful life) specifications. 

 

 

 

Step 12 – IVL Catheter Replacement 

Detach the IVL Catheter by first sliding the sterile sleeve out of the way moving it proximally along the IVL Connector Cable. Next, gently pull 

the CATHETER CONNECTOR and IVL Connector Cable apart to separate the IVL Catheter from the IVL Connector Cable connector (see 

illustration). Take care not to contaminate the connectors with fluids or other foreign matter during this procedure.  Position IVL Connector 

Cable to help ensure the connector remains free of contamination until the IVL Catheter can be replaced.  

   

CAUTION: Discard the used IVL Catheter per standard hospital procedures.  Used IVL Catheters cannot be re-sterilized and are designed for 

single use only.  Re-use of IVL Catheters can lead to patient injury. Connect a new IVL Catheter and resume patient treatment following the 

steps outlined above, beginning at step #5. Refer to the IVL Catheter IFU for information about recommended balloon overlap to prevent 

geographic miss. However, care must be taken not to exceed the recommended maximum number of pulses in the same treatment segment as 

indicated in the IVL Catheter IFU. 

 

 

 

CAUTION 

EQUIPMENT DAMAGE 

The IVL Generator delivers low-energy, short-duration, high-voltage pulses to the IVL Catheter through the IVL Connector Cable.  The system is designed not to deliver 

pulses unless an IVL Catheter connector is mated with the IVL Connector Cable and IVL Generator.  It is important not to allow the contacts or internal surfaces of 

unmated connectors to be contaminated with fluids.  Do not allow any connector to become contaminated by or immersed in fluids.  Failure to observe these 

precautions may damage the cables or catheters. 

 

4.1

 

Additional Usage Information 

The following topics and additional information may be useful in the use of the IVL Generator in treatment scenarios that may differ from the basic sequence of events defined 

in Section 4.0 above. 

 

TOPIC 

ADDITIONAL INFORMATION 

Tones 

The IVL Generator is designed to supplement its visual indicators with tones.  The IVL Generator will sound tones as follows: 

 

Click – Sounds once per therapy pulse to confirm therapy delivery in process. 

 

Two beeps – Positive confirmation of a user action.  Occurs when connecting a valid catheter or when arming or disarming the IVL Generator.  Also occurs at 

the end of the pre-programmed therapy pause period. 

 

Three beeps – Negative prompt.  Occurs when attempting to activate the IVL Generator but when one or more conditions are preventing activation.  Examples 

include attempting to activate the IVL Generator without a valid IVL Catheter attached or while holding down the THERAPY button.  Also occurs when the IVL 

Generator is deactivated and in the event built in diagnostics detect a malfunction. 

Use of 

Multiple IVL 

Catheters 

 

Multiple IVL Catheters may be used and re-used during the treatment of a single patient.  The IVL Generator is designed to track the useful remaining life of 

each IVL Catheter; however, no more than one of any given size IVL Catheter may be in use simultaneously.   

 

Connect and use IVL Catheters of varying sizes by following steps 5 – 9 above. 

 

Discard used IVL Catheters after patient treatment.  Used IVL Catheters cannot be re-sterilized and are for single use only.  Re-use of IVL Catheters can lead to 

patient injury. 

Automatic 

Off Feature 

The IVL Generator is designed to turn itself off and conserve battery power after periods of inactivity as follows: 

 

No IVL Catheter is connected – Will turn off after five hours. 

 

IVL Catheter connected – Will turn off after one hour. 

 

If the IVL Generator has turned itself off, simply press the POWER ON/OFF button to turn the IVL Generator back on.  Patient treatment may resume by 

following steps outlined in Section 4.0.  

Following 

Usage 

Follow these steps to prepare the IVL Generator for future use: 

 

Press POWER ON/OFF button once to turn the IVL Generator off. 

 

Remove and discard the IVL Catheter and sterile sleeve. 

 

Coil and secure the IVL Connector Cable 

 

Slide the CONNECTOR DOOR to the right to protect the THERAPY CONNECTOR 

 

If the last case of the day has been completed, or if the BATTERY CHARGING symbol is indicating empty, then place the battery on charge. The Charger Module 

must be connected to AC Mains and to the IVL Generator CHARGER CONNECTOR per Sections 5.2 - 5.3. 

 

Confirm the IVL Generator is charging; a lightning bolt symbol should appear inside the battery symbol. 

NOTE: The battery is self-discharging and requires periodic recharges even during storage so that the battery will not discharge to an unacceptably low voltage level, which 

could ultimately result in battery damage. 

 

 

 

 

 

Содержание IVL

Страница 1: ...ion and Test 9 6 Maintenance 10 6 1 Daily Maintenance 10 6 1 1 Charging and Testing the Internal Battery 10 6 1 2 Testing the IVL Generator 11 6 1 3 Inspecting the IVL Generator 11 6 1 4 Cleaning the IVL Generator 11 6 2 Monthly Maintenance 11 6 3 Other Maintenance 12 6 4 Product Useful Life 12 7 Troubleshooting 12 8 Appendix A Electromagnetic Compatibility Guidance 13 9 Appendix B Symbols 14 10 A...

Страница 2: ...result in serious personal injury or death CAUTION Hazards or unsafe practices that may result in minor personal injury product damage or property damage 2 2 Contraindications There are no specific contraindications for use of the IVL Generator and Connector Cable However users should read and understand the specific indications contraindications warnings and precautions included with the applicab...

Страница 3: ...rature and humidity conditions for at least twenty four hours before use See Appendix C for specified operating conditions Operating the equipment outside of these environmental conditions may cause equipment malfunction or damage IMPROPER DEVICE PERFORMANCE HAZARDS Using other manufacturers cables catheters power adapters or batteries may cause the device to perform improperly and may invalidate ...

Страница 4: ...ailable Counts down from available Pulse Count Per Catheter during treatment as each pulse is delivered Refer to applicable IVL Catheter IFU for the Max Pulse Count THERAPY STATUS Indicator On green when the device is ready to deliver therapy Flashes to indicate therapy is in process On yellow when therapy is paused or deactivated See Section 4 Steps 5 9 3 3 Front Panel Connectors FEATURE DESCRIPT...

Страница 5: ...NECTOR end of the Connector Cable to the THERAPY CONNECTOR Orient the connector and gently push in The connector is magnetic and will engage as the magnet gets close Push gently to confirm connector is fully engaged Slide CONNECTOR DOOR to left to reveal the THERAPY CONNECTOR THERAPY CONNECTOR Step 5 Prepare IVL Catheter for Use Prepare the catheter for use following the instructions in the IVL Ca...

Страница 6: ...nt as indicated in the IVL Catheter IFU CAUTION EQUIPMENT DAMAGE The IVL Generator delivers low energy short duration high voltage pulses to the IVL Catheter through the IVL Connector Cable The system is designed not to deliver pulses unless an IVL Catheter connector is mated with the IVL Connector Cable and IVL Generator It is important not to allow the contacts or internal surfaces of unmated co...

Страница 7: ...unting the IVL Generator to an unstable IV pole could present a tip over hazard to staff or patients 5 1 IV Pole Mounting STEP 1 Locate and identify mounting hardware see image below Two identical sets of IV pole mounting hardware are provided One set is used to mount the IVL Generator and one set is used to mount the battery Charger Module IV Pole Mounting Hardware STEP 2 Determine the IV pole di...

Страница 8: ...econd bracket so that it rests on top of the first bracket Orient the brackets so that they are facing opposite directions STEP 4 Mount the IVL Generator to the top bracket Align the holes in the mounting plate on the right side of the IVL Generator with the bracket mounting studs Push the IVL Generator towards the bracket to engage the mounting studs then push the IVL Generator down to lock in pl...

Страница 9: ...orage so that the battery will not discharge to an unacceptably low voltage level which could ultimately result in battery damage 5 4 Environment Conditioning The IVL Generator is designed to be used indoors in a controlled environment Refer to Appendix C for specified operating conditions Allow the IVL Generator to be stored in the ambient conditions of the usage environment for at least 24 hours...

Страница 10: ...IVL Generator Spilled liquids may cause the IVL Generator to perform inaccurately or malfunction 6 Maintenance This section describes the maintenance that clinical and or biomedical staff should be familiar with performing on a regular basis Recommendations for regular maintenance and maintenance intervals are noted below NOTE The battery is self discharging and requires periodic recharges even du...

Страница 11: ... this button for three seconds is required Step 4 Confirm Output Test Result The output test requires approximately 15 seconds to complete During this time the THERAPY STATUS indicator will remain on green Upon successful completion of this test the IVL Generator will sound four beeps If no error message appears on the display then this test has passed successfully 6 1 3 Inspecting the IVL Generat...

Страница 12: ...required Low BATTERY CAPACITY indicated when the battery symbol is Empty Battery requires charge Connect Charger Module to the IVL Generator and to AC Mains Allow the IVL Generator to charge at least twelve hours prior to use System Error displayed and red light around POWER ON OFF button Built in tests have detected a malfunction within the IVL Generator Turn IVL Generator off wait one second and...

Страница 13: ... hospital environment If the user of the Generator Model 825DP requires continued operation during power mains interruptions it is recommended that the Generator Model 825DP be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typic...

Страница 14: ...affected by absorption and reflection from structures objects and people Essential Performance Generator Model 825DP maintains safe and effective performance in the delivery of IVL therapy when operated in the electromagnetic environment specified in the table above 9 Appendix B Symbols The IVL Generator bears the following symbols Symbol Description Symbol Description Refer to instruction manual ...

Страница 15: ...stance 10 mL saline from above Generator 100 mL saline from any angle Connector Cable distal end Weight 15 pounds 6 8 kg 10 2 Appendix C2 Performance Specifications This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP Specifications Performance Characteristics Battery Rechargeable Smart Lithium Ion Battery Pack 14 4V 6 6Ah Ch...

Страница 16: ...PN 62393 Rev A Manufactured by Shockwave Medical Inc 48501 Warm Springs Blvd Suite 108 Fremont CA 94539 USA www shockwavemedical com patents MedPass SAS 95 bis Boulevard Pereire 75017 Paris France ...

Отзывы: