LBL 61876 rev. E, March 2018  

EN 

15 

 

10.

 

Appendix C: Specifications 

This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP.  All specifications are typical at 20°C unless 

otherwise stated. 

10.1

 

Appendix C1: General Specifications 

Specifications 

Performance Characteristics 

Alarms 

Built in tests and monitors are designed to detect and annunciate designated malfunctions of subsystems within the Generator.  The Generator is 

designed to interrupt therapy delivery in the event a malfunction is detected.  In addition, visual annunciators (CATH or SYS) will activate and three 

beeps will sound.  See Sections 4.1 (Tones), 7.0 (Troubleshooting) 

Classification, Product 

Class II Medical Electrical (ME) Equipment 

Classification, Applied Parts 

Type CF 

Connectors (Connector 

Cable) 

OnAnOn 150PT series with proprietary keyway 

Data Log 

No data associated with patient cases is logged 

Enclosure 

Non-ventilated, polymeric enclosure, molded from flame retardant UL 94V-0 rated material 

Environmental 

Altitude: 0 - 2000 meters  

Humidity, operating: 10 – 70% non-condensing 

Temperature, operating: 10° to 35°C 

Temperature, storage / shipping: -20° to 65°C 

Electrical Safety 

ISO 60601-1 (2012) edition 3.1 standards 

EMC Compatibility  

See Appendix A 

Mobility 

Product is designed to be mounted to a stable mobile or stationary IV pole.  An IV pole with five casters located in a circular pattern with a diameter of 

at least 23 inches (58 cm), such as the I.V. League Ventilator Stat-Stand™ model 1059 (or equivalent) is recommended.  

Power 

110 – 240 VAC; 50-60Hz; Single Phase, 15A service 

Protectively earthed 

Size 

11” (28.0 cm) high x 6” (15.2 cm) wide x 11.5” (29.2 cm)deep 

Shock 

Shipping Shock per EXD-007C ASTM D 4169-09 

Splash Resistance 

10 mL saline from above (Generator) 

100 mL saline from any angle (Connector Cable distal end) 

Weight 

15 pounds (6.8 kg) 

 

10.2

 

Appendix C2: Performance Specifications 

This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP. 

Specifications 

Performance Characteristics 

Battery 

Rechargeable Smart Lithium Ion Battery Pack (14.4V, 6.6Ah) 

Charge time less than twelve hours to full charge 

Fully charged battery capacity 12 patient cases 

(patient case: 900 therapy pulses delivered over one hour) 

80% capacity after 300 full charge/discharge cycles 

Meets requirements for transport via commercial aircraft  

(UN Manual of Tests and Criteria part III subsection 38.3) 

Diagnostics 

Built in tests and monitors are designed to detect and annunciate designated malfunctions of subsystems within the IVL Generator.  The IVL Generator 

is designed to interrupt therapy delivery in the event a malfunction is detected. 

Emitter Drive Channels 

Four channels, one to four channels may be used depending on catheter model connected. 

Output 

Proprietary pulse delivery system.  Output voltage 1000 – 3000 volts peak to peak, pulse duration ~ 1μS, pulse frequency 1, 2 or 4 Hz depending on 

catheter model connected. 

Output Voltage Accuracy 

The open circuit voltage at the IVL Generator THERAPY CONNECTOR: 5% of pre-programmed set point. 

Output Limits 

IVL System is designed to override user input and limit the number of continuous pulses delivered based on the IVL Catheter model connected.  Refer to 

IVL Catheter IFU. 

Therapy Settings 

Proprietary pulse delivery system.  No user adjustable settings.  Pulse delivery settings are pre-programmed, based on IVL Catheter model connected.  

Settings and IVL Catheter model detection employ redundant features.   

 

 

 

 

 

Shockwave Medical, Inc. 

48501 Warm Springs Blvd., Suite 108 

Fremont, CA 94539, USA 

www.shockwavemedical.com 

 

 

 

 

 

MedPass SAS 

95 bis Boulevard Pereire 

75017 Paris 

France 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание IVL

Страница 1: ...ion and Test 9 6 Maintenance 10 6 1 Daily Maintenance 10 6 1 1 Charging and Testing the Internal Battery 10 6 1 2 Testing the IVL Generator 11 6 1 3 Inspecting the IVL Generator 11 6 1 4 Cleaning the IVL Generator 11 6 2 Monthly Maintenance 11 6 3 Other Maintenance 12 6 4 Product Useful Life 12 7 Troubleshooting 12 8 Appendix A Electromagnetic Compatibility Guidance 13 9 Appendix B Symbols 14 10 A...

Страница 2: ...result in serious personal injury or death CAUTION Hazards or unsafe practices that may result in minor personal injury product damage or property damage 2 2 Contraindications There are no specific contraindications for use of the IVL Generator and Connector Cable However users should read and understand the specific indications contraindications warnings and precautions included with the applicab...

Страница 3: ...rature and humidity conditions for at least twenty four hours before use See Appendix C for specified operating conditions Operating the equipment outside of these environmental conditions may cause equipment malfunction or damage IMPROPER DEVICE PERFORMANCE HAZARDS Using other manufacturers cables catheters power adapters or batteries may cause the device to perform improperly and may invalidate ...

Страница 4: ...ailable Counts down from available Pulse Count Per Catheter during treatment as each pulse is delivered Refer to applicable IVL Catheter IFU for the Max Pulse Count THERAPY STATUS Indicator On green when the device is ready to deliver therapy Flashes to indicate therapy is in process On yellow when therapy is paused or deactivated See Section 4 Steps 5 9 3 3 Front Panel Connectors FEATURE DESCRIPT...

Страница 5: ...NECTOR end of the Connector Cable to the THERAPY CONNECTOR Orient the connector and gently push in The connector is magnetic and will engage as the magnet gets close Push gently to confirm connector is fully engaged Slide CONNECTOR DOOR to left to reveal the THERAPY CONNECTOR THERAPY CONNECTOR Step 5 Prepare IVL Catheter for Use Prepare the catheter for use following the instructions in the IVL Ca...

Страница 6: ...nt as indicated in the IVL Catheter IFU CAUTION EQUIPMENT DAMAGE The IVL Generator delivers low energy short duration high voltage pulses to the IVL Catheter through the IVL Connector Cable The system is designed not to deliver pulses unless an IVL Catheter connector is mated with the IVL Connector Cable and IVL Generator It is important not to allow the contacts or internal surfaces of unmated co...

Страница 7: ...unting the IVL Generator to an unstable IV pole could present a tip over hazard to staff or patients 5 1 IV Pole Mounting STEP 1 Locate and identify mounting hardware see image below Two identical sets of IV pole mounting hardware are provided One set is used to mount the IVL Generator and one set is used to mount the battery Charger Module IV Pole Mounting Hardware STEP 2 Determine the IV pole di...

Страница 8: ...econd bracket so that it rests on top of the first bracket Orient the brackets so that they are facing opposite directions STEP 4 Mount the IVL Generator to the top bracket Align the holes in the mounting plate on the right side of the IVL Generator with the bracket mounting studs Push the IVL Generator towards the bracket to engage the mounting studs then push the IVL Generator down to lock in pl...

Страница 9: ...orage so that the battery will not discharge to an unacceptably low voltage level which could ultimately result in battery damage 5 4 Environment Conditioning The IVL Generator is designed to be used indoors in a controlled environment Refer to Appendix C for specified operating conditions Allow the IVL Generator to be stored in the ambient conditions of the usage environment for at least 24 hours...

Страница 10: ...IVL Generator Spilled liquids may cause the IVL Generator to perform inaccurately or malfunction 6 Maintenance This section describes the maintenance that clinical and or biomedical staff should be familiar with performing on a regular basis Recommendations for regular maintenance and maintenance intervals are noted below NOTE The battery is self discharging and requires periodic recharges even du...

Страница 11: ... this button for three seconds is required Step 4 Confirm Output Test Result The output test requires approximately 15 seconds to complete During this time the THERAPY STATUS indicator will remain on green Upon successful completion of this test the IVL Generator will sound four beeps If no error message appears on the display then this test has passed successfully 6 1 3 Inspecting the IVL Generat...

Страница 12: ...required Low BATTERY CAPACITY indicated when the battery symbol is Empty Battery requires charge Connect Charger Module to the IVL Generator and to AC Mains Allow the IVL Generator to charge at least twelve hours prior to use System Error displayed and red light around POWER ON OFF button Built in tests have detected a malfunction within the IVL Generator Turn IVL Generator off wait one second and...

Страница 13: ... hospital environment If the user of the Generator Model 825DP requires continued operation during power mains interruptions it is recommended that the Generator Model 825DP be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typic...

Страница 14: ...affected by absorption and reflection from structures objects and people Essential Performance Generator Model 825DP maintains safe and effective performance in the delivery of IVL therapy when operated in the electromagnetic environment specified in the table above 9 Appendix B Symbols The IVL Generator bears the following symbols Symbol Description Symbol Description Refer to instruction manual ...

Страница 15: ...stance 10 mL saline from above Generator 100 mL saline from any angle Connector Cable distal end Weight 15 pounds 6 8 kg 10 2 Appendix C2 Performance Specifications This appendix contains the specifications and performance characteristics for the Shockwave Medical IVL Generator Model 825DP Specifications Performance Characteristics Battery Rechargeable Smart Lithium Ion Battery Pack 14 4V 6 6Ah Ch...

Страница 16: ...PN 62393 Rev A Manufactured by Shockwave Medical Inc 48501 Warm Springs Blvd Suite 108 Fremont CA 94539 USA www shockwavemedical com patents MedPass SAS 95 bis Boulevard Pereire 75017 Paris France ...

Отзывы: