17
4.4
Interface
Displayed when [
] is [
~
].
Item
Set Value
Description
Output Control
Stop output
Output continuously at all times
Output continuously if stable (Stop output if unstable)
Output once when the Print key is pressed (regardless of whether
the balance is stable or unstable).
Output once when the balance is stable. After any sample is lifted
causing the display to show zero or less value, outputs the next
time any sample is put on again and the balance is stabilized.
Output once when the balance is stable. Stop output when
unstable. Outputs once when the balance is stabilized again
(including the zero point) even if it is not reloaded.
Output once when the balance is stable. Output continuously
when unstable. Output is stopped after a single output when the
balance is stable even if it is not reloaded.
Output once when the Print key is pressed if the balance is stable.
Output once immediately after a specified interval.
Output once immediately after a specified interval if the balance is
stable.
Baud Rate
1200 bps
2400 bps
4800 bps
9600 bps
19200 bps
Parity
None
Displayed when [
] is set to [
to
].
Odd
Even
Data Length
7 bits
Displayed when [
] is set to [
,
].
8 bits
Stop Bits
1 bit
2 bits
Unused Higher Digits
Embed 0 (30H).
Filled by space characters (20H).
Response Commands
The A00 and Exx responses are used.
The ACK and NAK responses are used.
A star ( ) denotes a factory setting.
* The data output interval is 0.1 to 1 second for continuous output.
(The interval varies with the measurement conditions, etc.)