Shining 3D EinScan SE/SP_User Manual_3.1.1
14
3.
Calibrate the Scanner
3.1.
Precautions and Use
Calibration is the process to ensure the device will scan with the optimal accuracy and scan quality.
Each EinScan 3d scanner was factory-tested and achieved, at least, the listed accuracy from the
technical specification according to VDI/VDE 2634 standards.
Besides factory tested guarantee of accuracy, to assure high scanning accuracy, each EinScan is also
coming with a calibration board for customer calibration. This customer calibration is not required
daily often. Only when the first time after scanner and software are installed. Or, when you are in the
following situations, you should calibrate the device:
When the scanner is used for the first time or after long time without using.
When there is strong vibration during the transportation.
When alignment mistake or failure frequent appear during the scanning.
When scanning data is incomplete and quality is much worse during the scanning.
3.2.
Operation
After installation, when you open the software for the first time, choose device type and it will enter
the calibration interface automatically. You can also choose Calibration on the navigation bar to enter
calibration process later.
If there is no calibration data, click “quit”, the software will prompt “No
calibration data, please calibrate”.
Note:
1.
Calibrate the Scanner twice a week for accurate scanning.
2.
Make sure to protect the calibration board and keep it clean, no scratches or stains on the black
surface with white circles.
3.
The Calibration board is matched to the Device with same Serial Number. Doing the calibration
with an incorrect calibration board will fail to generate good scan data or optimum accuracy.
4.
Clean with clear water only, do not use alcohol or chemical liquid to clean the calibration board