Shin-Nippon SLM-5000 Скачать руководство пользователя страница 18

 

RB-473-B02M 

17 

5.4. Printer 

5.4.1.  How to operate 

When measurement is finished, press              . The measurement result is printed out. 

 

5.4.2.  Type of print out 

<<Print out sample when lenses for right and left eyes are measured>> 

Store name, comments, etc
(printed out when registered only)

Lens data for right eye

Lens data for left eye

Line feed

<R>

S : 

-

0.00

C : -2.75

A :      90

ADD:+3.00

P : O 1.50 U 1.50

<L>

S : 

-

0.00

C : -3.00

A :     

-

0

ADD: +3.00

P : O 0.00 D 3.50

SHIN-NIPPON      SLM-5000

2002 03 25                          10:25

The unit of the prism value

depends on the setting.

NAME

 

 

<<Print out sample when a single focal lens is measured>> 

 

ADD  value  is  printed  out  when
mutifocal  or  progressive  lens  is
measured.

F   V

N   V

Optical characteristics graph printed.
(When Graph Print is set for On at the
progressive lens measurement.)

< S >

A D D : + 2 . 0 0

P :   O   0 . 0 0   U   3 . 2 5

SHIN-NIPPON            SLM-5000

2 0 0 2   0 8   2 5             1 0 : 2 5

N A M E

< S >

S   :   - 3 . 0 0
C   :   - 0 . 5 0

A   :       9 0

P   :     0 . 0 0     2 5 3   B

N A M E

SHIN-NIPPON            SLM-5000

SHIN-NIPPON            SLM-5000

< S >

A D D :   + 2 . 5 0

P   :   O   1 . 2 5   U   5 . 0 0

N A M E

2 0 0 2   0 8   2 5             1 0 : 2 5

2 0 0 2   0 8   2 5             1 0 : 2 5

S   :  

-

0 . 0 0

C   :   - 2 . 7 5

A   :       9 0

S   :   + 2 . 2 5
C   :  

-

0 . 0 0

A   :         0

 

Содержание SLM-5000

Страница 1: ...RB 473 B02M Operations Manual Important Read this manual thoroughly before use Keep this manual on hand at all times Auto Lensmeter SLM 5000...

Страница 2: ...NOTE 1 The information contained in this manual is subject to change without notice 2 While reasonable efforts have been made in the preparation of this document to ensure its accuracy you should con...

Страница 3: ...ny liquid gets on the instrument or when any foreign matters gets inside the instrument unplug the power cord immediately and contact your local distributor In case of occurrence of abnormality noise...

Страница 4: ...d clean without any dust or oil on it The dirty terminal may cause malfunction or fire When the power cord becomes hot after used check for the dirt at the terminal part If you find no dirt replace th...

Страница 5: ...t of marking pen 16 5 4 Printer 17 5 4 1 How to operate 17 5 4 2 Type of print out 17 5 4 3 Setting and replacement of printer paper 18 6 Measurement 19 6 1 Check before Measurement 19 6 2 Single Lens...

Страница 6: ...Parts Identification 1 1 Overview Printer Power switch Power inlet RS 232C terminal Rating nameplate Color LCD monitor Operation switches Indicator Marking lever Lens plate Lens stand Lens holder Mem...

Страница 7: ...heck the followings are all packed without any damages Contact your distributor if any accessories are missing or damaged Power cord Contact lens stand Rolls of printer paper Printer shaft Dustproof c...

Страница 8: ...re range for use 10 C 40 C Temperature range for storage 10 C 60 C Humidity range for use storage 70 HR no dew condensation allowed 4 Keep away from inflammable or explosive gases as well as storage a...

Страница 9: ...ith care at all times making sure not to accidentally drop it 5 If the instrument is not to be used for any length of time remove the power cord from the outlet 6 When not in use the instrument should...

Страница 10: ...the addition is measured determines the far point for auto measurement of the progressive lens For the manual measurement determines the measurement positions and values for the far and near points ME...

Страница 11: ...r near point after the progressive lens is measured Off not printed On printed Off Prism Selects whether to display prism values and which unit is used if displayed Off not displayed X Y X Y display P...

Страница 12: ...itch to call up Screen 2 Using and move the cursor to the letter or symbol you wish to change Then release Memory Add switch to return to the Screen 1 Select the letter or symbol you wish to change an...

Страница 13: ...te Transfer rate 38400 NB Bit Character length data bit 8 fixed Parity Parity check None fixed Stop Bit Stop bit 1 fixed Signal names and connection Connect each connector of the serial interface as s...

Страница 14: ...M D Y year date month Date change of the date Move the cursor to the item you wish to change with While Memory Add switch is pressed Screen 2 will be displayed Make changes using Time change of the ti...

Страница 15: ...ush the PUSH mark on the lens plate to unlock it The lens plate will come out To lock the lens plate push it again 2 The lens plate comes out until it touches the lens when pushed Place the frame lens...

Страница 16: ...nger slowly after finishing marking on a lens 4 Marking lever returns to the initial position Observe the following otherwise you may damage the pen tip Do not push the marking pen too hard when marki...

Страница 17: ...en as shown below Then insert the pen back to the initial position 3 Holding the marking pen fix it with the screw note A central marking pen is black and is shorter than pens on both sides Use the ma...

Страница 18: ...O 0 00 D 3 50 SHIN NIPPON SLM 5000 2002 03 25 10 25 The unit of the prism value depends on the setting NAME Print out sample when a single focal lens is measured ADD value is printed out when mutifoc...

Страница 19: ...sert the printer shaft into the roll of the paper and set it in the printer unit Printer shaft Printer paper 3 Hook the tabs of the cover to the tabs of the unit Take out the end of the paper a little...

Страница 20: ...5 4 3 Setting and replacement of Printer paper Check that no lenses are placed on the lens stand Turn the power on The screen below will appear in a few seconds Always make sure that the cable is gro...

Страница 21: ...nsmeter and the cross cursor that of the lens 3 Move the lens until the cross cursor and the alignment matches When they match the message Marking OK will appear to indicate that marking is ready to b...

Страница 22: ...y press Press to print out the values Great care should be taken not to give strong impact to the lens when lowering the lens holder Confirm that the lens holder is locked at the top when pushing it u...

Страница 23: ...lues stored in memory are shown in red After stored in the memory the values displayed on the screen will be fixed 5 Press to switch to the lens for a left eye At this time the measurement result for...

Страница 24: ...point in memory After the measurement values are stored ADD values will become red For a trifocal lens Press Memory Add switch again to display Ad2 on the screen Bring the second near point into the c...

Страница 25: ...placed on the lens stand automatic judgment may also fail In such a case try to move the lens slowly back and forth or to the right or left When ADD value is small under 1D the lens in the frames is...

Страница 26: ...s green Then the measurement value of the far point is stored in memory 3 Near point measurement Move the lens slowly toward you according to the arrow as shown in the screen 4 so that the alignment c...

Страница 27: ...o so In such a case measure manually based on the graphs of Add value and assessment To measure the far point manually carry out alignment a ccording to the shape of the alignment cursor as you do for...

Страница 28: ...asurement value is stored automatically in memory when the alignment mark and the cross cursor overlap as shown on the left 6 7 Power saving function Power saving function is activated when no switche...

Страница 29: ...in the contact lens stand as shown below 2 Remove the standard lens stand and mount the contact lens stand 3 Lower the lens holder and hold the contact lens stand 4 Move the contact lens stand vertica...

Страница 30: ...e target so that the AXIS mark matches the value in the prescription 2 To be more precise align according to the AXIS value measured Marking on the cylindrical axis 1 Match the target so that the AXIS...

Страница 31: ...displayed on the screen accord with those of the prescription P Prism value B Base direction Prescription expressed in mm 1 Select On from Prism mm in the Menu screen 2 Move the lens so that the prism...

Страница 32: ...lens stand when the power is turned on or if the lens under the lens stand is dirty Remove the lens if it is on the lens stand If the lens under the lens stand is dirty gently wipe it with a soft clo...

Страница 33: ...ens Soft contact lens A special accessory is required Power supply 100 240V 50 60Hz Printer Thermal printer paper width 58mm Monitor Color LCD monitor 5 7 inches Dimension weight 200 W x260 D x443 H a...

Страница 34: ...ve Distributor Manufacturer RyuSyo Industrial Co Ltd Kagawa Factory 958 IKEUCHI KOUNAN CHO TAKAMATSU KAGAWA 761 1494 JAPAN AJINOMOTO TRADING INC SHIN NIPPON Medical Ophthalmic Instruments Dept EAST WI...

Отзывы: