12
Com mun ica t i on Wi t h Ex t e r n a l Dev ices (con t i n ued )
External Control Command Descriptions (continued)
APF ____________________ Auto-Power-Off
APF CR = 0: none
3: 3 minutes
EXS ________________________External Input Setting
EXS CR = Z: TARE
H: HOLD
N: NONE
HLD _________________ HOLD Trigger Mode
HLD CR = E: EDGE
L: LEVEL
O: Open Contact
C: Closed Contact
LoG __________________ Memory Mode Setting
LoG CR = S: Continuous
N: Standard
PKF __________________ PEAK Fast Mode
PKF CR = 0: OFF
1: ON
DAF __________________ Analog Output Setting
DAF CR = 0: 1 msec
1: Display update
ADF __________________ Filter Setting
ADF CR = 0: None filter
1: Normal filter
2: High filter
D, DATN, DATT, DATC, AB, BB, BC, BD _______Data
output request (8 commands)
D
CR ....Output displayed DATA
DATN
CR ....Output average mode DATA
DATT
CR ....Output tension PEAK DATA
DATC
CR ....Output compression PEAK DATA
AB
CR ....Stop continuous data output
BB
CR ....Continuous data output
BC
CR ....Model name confirmation request
BD
CR ....Unit's confirmation request
Typical
gauge
response
LIST ____________ Gauge’s present state
LIST
CR (Command from PC)
MDL
DRI-500 lb
Units
lb
DSPM
NORMAL
DSPT
0.125
PKFST
ON
DAFST
ON
ADFIL
NONE
OFFT
NONE
232C
4800
B8
S2
232C
PN
CR LF
X OFF
CP
ON
HLMT
50.00
LLMT
- 10.00
ET
HOLD
ET
LEVEL OPEN
LOG
CONTINUE
Gauge Response
OK
....Command accepted
ERRO
....Did not recognize command*
ERR1
....Command cannot be processed*
ERR2
....Wrong nomenclature, i.e N,S,C*
ERR3
....Communications error*
ERR7
....Memory unit error**
ERR8
....Sensor error**
IN HOLD
....When in HOLD, certain commands
cannot be processed.
IN MEMORY
....When in continuous memory mode or
standard mode (active), certain
commands can be processed.
* Check command and resend
** See troubleshooting section
UNITS CHANGE DISABLE .... This output is transmitted to
PC when someone is trying to change UNITS while memory is
accepting DATA.