Shimpo Instruments DT-326B Скачать руководство пользователя страница 6

Trigger Edge Setting (External Modes)
In external mode, the trigger edge parameter will allow for select-
ing either the leading edge or trailing edge of the input pulse to 
trigger the stroboscope flash.

In parameter mode, use the SET key to cycle to the Trigger Edge 
setting parameter mode. To set the trigger edge as “Down Edge” 
(trailing), press “ -“ key or turn the dial clockwise. To set the trig-
ger edge as “Up Edge” (leading), press “+” key or turn the dial 
clockwise.

When trigger edge setting is set to “up edge”. The flash will oc-
cur on the leading edge of the input pulse (when delay setting is 
zero).

External Pulse Signal from Machine/Sensor 

vs. Stroboscope Flash

Up Edge

When trigger edge setting is set to “down edge”. The flash will 
occur on the trailing edge of the input pulse (when delay setting 
is zero).

External Pulse Signal from Machine/Sensor 

vs. Stroboscope Flash

Down Edge

Back Light Setting
The LCD display of the DT-326B stroboscope has a backlight that 
can be turned on or off in the parameter settings. Turning off the 
backlight will conserve battery power. To adjust the backlight set-
ting, enter parameter mode and use the “SET” key to cycle to LCD 
setting. Press “+” key to turn on the backlight. Press “-” key to 
turn off the backlight.

Note: If the backlight is set to on, as a power saving feature, it will 
automatically shut off after the set minutes if no keypad or dial 
operation has occurred. The LCD backlight will turn back on with 
keypad or dial operation.

Automatic Power-Saving Settings 

Auto Power Off: The unit will automatically turn off after 1 to 
1000 minutes depending on settings if no keypad or dial oper-
ation occurs when using battery power. Flashing will resume 
with either a dial (rotation) movement or a keypad operation.

The LED auto shutoff will not occur in External trigger mode, 
while a signal is being received by the unit. The LED flashing 
will automatically turn off after the set 2 minutes of no external 
pulse input.

LCD Back Light Off: If the backlight is set to on, as a power 
saving feature, it can be adjusted to automatically shut off from 
1 to 1000 seconds, if no keypad or dial operation has occurred. 
The LCD backlight will turn back on with keypad or dial oper-
ation.

Factory Reset (Initial)
To reset unit to factory values choose “Yes” under Initial in the 
parameter setting mode. Then press “SET”.

7