Shimpo Instruments DT-326B Скачать руководство пользователя страница 2

LCD Display

Main Data Display

•Flash rate value will be displayed in internal flashing or Exter-
nal trigger mode.

•Display will indicate “P” or “LCD” (in ON/OFF setting of LCD
backlight) in *Parameter setting mode.

*For details about Parameter setting mode, please refer to Pa-
rameter setting mode.

Sub Data Display

Internal Mode
•Phase Shift will indicate degrees of phase shift in internal flash-
ing mode. To change the degrees of phase shift, press + or - 
keys. The display will reflect the cumulative angle of phase shift.

•The flash duration time will be displayed when the RATIO key
is pressed. The flash duration time can then be increased or
decreased with the dial from 0 to 2.5 (0~2.5°) degrees.

External Mode
• In external trigger mode, the display will show the settings of
delay time. For details about changing these settings, please
see external trigger mode section.

Battery Indication 

- Indicates remaining battery charge.

Note: The display will not indicate a battery charge when the unit 
is being used with the AC Adapter.

Remarks: It can be used for more than 12 hours continuously 
when the frequency is 3000FPM or 50Hz.  

FUNCTION INSTRUCTIONS

Mode Selection - To switch between INTERNAL, EXTERNAL and 
PARAMETER mode, press and release the MODE key. For details 
about Parameter setting mode, please refer to Parameter setting 
mode section.

Internal Flashing Mode

Changing Units of Measure in Internal Mode - To change the 
measuring units from FPM to Hz, press and release the FPM/Hz 
key. Units will alternate with each press and release.

Flash Rate (Frequency Setting) in Internal Mode 
You can set the flash rate by turning the dial in the center of the 
unit. To increase the flash rate, turn the dial in the clockwise direc-
tion. To decrease the flash rate, turn the dial in the counter-clock-
wise direction. For small adjustments, turn the dial slowly. For 
quick adjustments, rotate the dial quickly. The flash rate range 
and resolution will be different according to the selected measur-
ing range. 

One function of a stroboscope is to provide a “stopped” image 
of a rotating target when the flash rate of the stroboscope has 
matched the rotational speed of the target object. The strobo-
scope will show a single image when the flash rate is set to a 
lesser multiple of the true RPM (1/2,1/3,etc.) When the flash rate 
is increased to a higher multiple (2,3,etc.), multiple images will 
appear. To find the true RPM of the target object, reduce the flash 
rate to a lower multiple until only one image appears. 

2

Battery
Indicator

Parameter
Icons

Sub
Data Display

Main

Data

Display

Units

Batter capacity is full

The Battery capacity is diminished Near zero 
capacity

When battery is empty, the unit will power off

1.800.561.8187 

[email protected]

-Direct.ca

Отзывы: