2. Connection of the system
2.1 Power transformers
•
A step down transformer, (XAT-2) is normally shipped together with this generator for installation.
This transformer can provide proper power input for UD150L-30 FX/EX generator, YSF/RS table
system, HMS/AID high-speed starter, and other peripheral accessories.
•
There are two other transformers in UD150L-30 FX/EX generator are power related, T1 and T2
transformer. It is necessary to change taps for these two transformers according to the supplied input
from XAT-2 before proceeding.
(1) T1 transformer,
For example, if the incoming power is three-phase 400V from XAT-2, connect AIN-0 to –A200 and
AIN-2 to A200.
(2) T2 transformer,
Connect T2-0 to –200 and T2-2 to 200 base on the same assumption as (1).
2.2 Cable connection
•
Power cable
Hospital breaker (3
φ
480V) XAT-2 transformer (3
φ
400V)
Control
cabinet
U
I/P
O/P
UN
Power
terminal
V
I/P
O/P
VN
W
I/P
O/P
WN
GND
GND
GND
GND
Terminal
block
Note: Only ground one end of the shielding on the three-phase power cable to prevent ground loop
from happening. Refer to Fig. 2.4.3 in the installation manual.
•
Control console
Control
console
Control
cabinet
J62
J
62
System
LF
30
PCB
J 63
J63
I/O LFX PCB
Frame
GND
Terminal strip
J9
Hand Switch
•
High speed starter (HMS or AID)
Connect JS1 cable between SYSTEM-LF 30 PCB and high speed starter by using JS1 connector kit in
the accessory package shipped with this generator.
Following table is the description of the signal used for high speed starter
Signal Meaning
Signal Meaning
2
Normal speed rotation
1M
Tube 1 selection
7
High speed rotation
2M
Tube 2 selection
H2
Normal speed confirmation
SBK
Starter fault signal
H7
High speed confirmation
CM1
Return for 7, H7, and SBK (0V DC)
F
Fluoro tube rotation
CM2
Return for 2 and H2 (+15V DC)
SP1
Spot film
CM3
Return for F and SP1 (L0)
CM4
Return for 1M and 2M (+15V DC)
UD150L30-FX/EX Installation Guide, REV_B Page 6 of 35