3.2.2 Power ON check up
•
Check all DC and AC power according to the table below.
Name Measuring
point
Voltage Note
T2 transformer
10LA-0LA
10.5V + 1V
100LB-0LB
105V
+
10V
T1 transformer
L100-L0
AC 105V + 5V
L125-L0
AC 130V + 10V
A125-A0
AC 125V + 5V
Without Load
A200-A0
AC 200V + 15V
135C-0C
AC 142V + 10V
20T-0T
AC 21V + 2V
DC power supply J73/J74 4-7
DC +5V + 50mV
PS1
J73/J74 2-5
DC +15V + 400mV
PS1
J73/J74 6-5
DC -15V + 400mV
PS1
J75 1-4
DC +24V + 600mV
PS2
•
Make sure MG-RS is activated. If not, turn power off immediately, and manually push the button on
MG-RS to see if that was a mechanical jam. Also, check the cable connection on it. Fire accident could
happen if this is not working.
•
Check to see if the charging voltage on VC+ and VC- (ANALOG-LF30 CP72 and CP73) is within
range. (+10%) (3
φ
400 V input power to generator will results the charging voltage on VC+ and VC-
approximately to be 560V) If not, you need to check the power line and XAT-2 again.
•
Perform IGBT Gate check as the following and make necessary adjustments before conducting X-ray.
(Please check this part after you finish initial settings)
Procedures:
1) Power off generator and wait about 5-10 seconds.
2) Measure VC+ and VC- with DVM to insure the charging voltage were all discharged.
3) Disconnect J202 and J203 on ANALOG-LF30 PCB to disable MG-L and MG-RS.
4) Add jumper between CP42 CVP1 and GND test point on ANALOG-LF30 PCB. Add another
jumper between the cathode side of D27 and CP58 +15V. (Connection Diagram ANALOG-
LF30 9/27) Add another jumper between CP52 CHECK1 and CP56 CHECK2 on ANALOG-
LF30 PCB.
5) Connect oscilloscope channel one to CP1 E and CP2 G on GATE DRIVER-99 PCB mounted
on IGBT-QA.(Probe signal on CP2, ground on CP1)
6) Connect scope channel two the same as channel one on IGBT-QC. (Float ground on channel
two probe to prevent ground loop)
7) Set amplitude 5V and time 5
µ
sec. per division for both channels on the scope. Place the
ground for both channels at the center of screen as well.
8) Turn on power of generator and select General Radiographic Mode. Then set 100 kV, 100
mA, and 2.5 sec.
9) Make an exposure and wait about 0.1 to 1 second until the X-ray operation reach to its steady
state.
10) Observe the waveforms and make the following adjustments if necessary.
11) Adjust VR12 PULSE ADJ on ANALOG-LF30 PCB so that the pulse width is 10.0 + 0.5
µ
sec. at 5 V level. Also, check the phase difference is 11.1 + 0.5
µ
sec. at 0 V level. The
positive peak voltage is less than 20 V and flat level is at 15 V or a little higher. The negative
flat level is about –6 V. Repeat step 9) and 11) several times for the best results you can
achieve. (Compare the waveform with the attached waveform.)
UD150L30-FX/EX Installation Guide, REV_B Page 15 of 35