TOCV-CP_IMshort_Control-ver-2_Rev07F
Page 12 of 22
(Software version ≥ 2.00)
effective date: September 2007
Overview about several possibilities to create a Calibration Curve:
Wizard -2
Calibration curve
type
Calibration points are
distributed uniformly over
concentration range
Edit calibration curve
point manually
Calibration curve
according to
DIN 38402/P-51
Calibration curve according to
USP/EP
Application
Normal calibration curve, data
points are added
automatically, helps by
multiple point calibration
curve
Normal calibration
curve, data points are
added manually
!Normally used!
Special calibration curve:
always10 points, first point
and last point are injected
10times all other points only
one time; Linearity and
homogeneity are calculated
Special calibration curve:
Always 2 points:
0 and 500ppb
(United States Pharmacopoeia or
European Pharmacopoeia
Wizard -3
Same
Same
Same
Same
Wizard -4
Units
Concentration range
No. of Cal points
No of Injections
No of Washes
SD Max
CV Max
Sparge time
Acid addition
Selectable default: ppm
Enter range: e.g 0-5
Enter No of Points: e.g.
3
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Selectable default: ppm
Disable
Disable
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Selectable default: ppm
Enter range: e.g. 10-100
Recommended: 10
Recommended: 1
Selectable (0-10)
Disable
Disable
Selectable (0-20min)
Selectable (0-20%)
Recommended: ppb
Fixed conc. range:0-500
Fixed No. of Cal. Points: 2
Selectable (1/20)
Selectable (0-10)
Selectable (0-9999)
Selectable (0-100%)
Selectable (0-20min)
Selectable (0-20%)
Wizard -5
Calib. Points list
Inj. Volume
Calibration Points
Recommended value
Example:
0ppm
5ppm
10ppm
Calculated automatic.
Empty calibration
points list. Manually
filling via [ADD]
Recommended value
List of 10 calibration points
in the range from 10 – 100
(as example)
Recommended value
List: 0ppb
500ppb
Wizard -6
Same
Same
Same
Same