7.5
Alarm Detection Capabilities
71
Power Supply Unit
INSTRUCTION MANUAL
7
Table 7-4
If the ALARM Lamp Lights
LCD Display
Possible Cause
Remedy
Secti
on
1
11=TD COUNTER
LIMIT
12=PF COUNTER
LIMIT
The number of high speed
or power failure touch-
downs has exceeded the
prescribed number.
The touch-down bearing may have
deteriorated. Consult Shimadzu or an
approved service company regarding
replacement of the touch-down bearing.
2
13=WRONG TMP
MODEL
The models of pump and
power supply do not
match.
Check the combination of the pump and
power supply models.
Check the connection of the magnetic
bearing cable. Also check that the
connector pins are not bent. Turn off the
power supply before checking the
magnetic bearing cable. Never disconnect
the cable while the power is on.
5.3
3
14=AC LOW
VOLTAGE
15=POWER FAILURE
Power failure or reduction
in the power supply
voltage.
Wait for the power to be restored. An
unfamiliar sound will be heard a few
minutes after a power failure. This sound
occurs when the rotor contacts the
protective bearing as magnetic levitation
cannot be maintained. This is not an
abnormal sound.
POWER switch was
turned off by mistake.
Wait about 5 seconds before turning the
POWER switch back on. Re-acceleration
is possible after resetting and start-up.
4
16=TMP:OVERLOAD
Drop in rotation speed
during rotation at rated
speed (increased internal
pump pressure).
Check that the outlet and inlet pressures
are below the specified maximum
pressures.
Check for leakage.
Check that too much purge gas is not
flowing.
Check that process gas flow rate is not too
high.
Содержание EI-D1003M
Страница 9: ...viii 263 13228 This page is intentionally left blank...
Страница 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Страница 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Страница 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Страница 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Страница 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Страница 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Страница 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Страница 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...
Страница 119: ...Appendix A COMMUNICATIONS A 28 263 13228 A This page is intentionally left blank...
Страница 121: ...Index Index 2 263 13228 Power Supply Unit INSTRUCTION MANUAL This page is intentionally left blank...