Appendix A
COMMUNICATIONS
A-14
263-13228
A
Table A-4
Table of Timer
Table A-5
Alarm History Data Format
No.
Name
Range
Description and format
01
Run time
00000 to 99999
Read the timer value in the “MENU MODE / INTEGRAL
TIMER / RUN TIME” on the LCD display. (Can not be
reset. Reset date is invalid.)
Example: 0
→
00000, 99999
→
99999
02
Last maintenance
time
00000 to 99999
Read the timer value in the “MENU MODE / INTEGRAL
TIMER / LAST MAINT.” on the LCD display. (Can be
reset)
Example: 0
→
00000, 99999
→
99999
03
Power failure touch-
down count
00000 to 00999
Read the timer value in the “MENU MODE / INTEGRAL
TIMER / POWER FAILURE” on the LCD display. (Can be
reset)
Example: 0
→
00000, 999
→
00999
04
High-speed touch-
down count
00000 to 00999
Read the timer value in the “MENU MODE / INTEGRAL
TIMER / MB ALARM” on the LCD display. (Can be reset)
Example: 0
→
00000, 999
→
00999
05
MB warning counter
00000 to 00999
Read the timer value in the “MENU MODE / INTEGRAL
TIMER / MB WARNING” on the LCD display. (Can be
reset)
Example: 0
→
00000, 999
→
00999
06
Maintenance call time
00000 to 99999
Read or change the timer value in the “MENU MODE /
INTEGRAL TIMER / MAINT.CALL” on the LCD display.
Example: 0
→
00000, 99999
→
99999
Item
Number
of bytes
Data
Comments
1
History number
2
01 to 99
History number designated by the command.
2
Time
10
YYMMDDHHMM
Time when the failure occurred (stored as
Greenwich Mean Time)
YY: year, MM: month, DD: day,
HH: hour, MM: minutes
3
Alarm number
2
00 to 99
Alarm number of the failure that occurred. See
Table 7-6 "Table of Alarms" and Table 7-7
"Table of Warnings" in this manual for details.
4
Run status
2
NS, NA, NN
⋅⋅⋅
Run status when the failure occurred. Data is
identical to CS command answer.
5
Rotational speed
4
0000 to 0100
Speed when the failure occurred. Format is
identical to 09 in Table A-3.
6
Motor current
4
0000 to 0150
Motor current in the event of a fault. The format
is the same as No.04 in Table A-3.
7
Pump temperature
2
00 to 99
Pump temperature in the event of a fault. The
format is equivalent to the last two characters of
No.05 in Table A-3.
(Valid only for a power supply with a temperature
control function.)
8
Temperature control
function
2
00, 01, 02
Temperature control function when the failure
occurred. Format is equivalent to last 2
characters of 07 in Table A-3.
Содержание EI-D1003M
Страница 9: ...viii 263 13228 This page is intentionally left blank...
Страница 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Страница 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Страница 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Страница 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Страница 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Страница 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Страница 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Страница 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...
Страница 119: ...Appendix A COMMUNICATIONS A 28 263 13228 A This page is intentionally left blank...
Страница 121: ...Index Index 2 263 13228 Power Supply Unit INSTRUCTION MANUAL This page is intentionally left blank...