74
6
ADJUSTING RESPONSE AND STABILITY
The stability mark is a symbol (
) that is displayed when it is determined that the weight reading has
stabilized.
The following settings adjust conditions for lighting up of
(the stability mark).
•
Stability detection range
•
Stability mark lighting timing
Normally there is no need to change these settings. (Change the settings if, for example, you want to relax
the conditions and make
(the stability mark) light more easily because the environment is unstable,
or to speed up operation if stability is used to automatically print or output data.
*
Lighting up of
(the stability mark)
The lighting up of
(the stability mark) indicates the fact that the weight reading is stable.
If the load is being changed slowly, or due to the settings relating to stability detection, the weight
reading may change while
(the stability mark) remains lit, or
(the stability mark) may light
temporarily and then the weight reading may change.
Setting the Stability Detection Range
The stability detection range is a value set as a count of the smallest digit that is displayed, and
the display is judged to be stable if fluctuation in the weight reading is within this count during a
fixed time.
(The default setting for the stability detection range is 1 count (1d).)
Effect of reducing the stability
detection range
It takes some time for
(the stability mark) to light up, but
after it has lit the weight reading is stable (improvement in
reliability).
Effect of increasing the stability
detection range
(the stability mark) can be made to light more quickly
but the weight reading is liable to fluctuate after it has lit
(improvement of weighing and data output speeds).
1
Press
twice in the weighing mode.
This opens the main menu.
2
Select setting of the stability
detection range.
[TOOLS]
[PARAM.W]
[BAND]
! ! !B!A!N!D
Adjusting the
Stability Mark
Содержание ATX-R Series