12) Setting of ramp unit (Group 2-14)
In this screen, second (Sec.) or minute (Min.) is selected as the unit of the rate of change.
Note 1: The unit of ramp time can be changed during execution of ramp control. When changed, ramp control is carried out
immediately on the new unit.
13) Setting of ramp rate (Group 2-14)
For the position of decimal point of PV display on set unit,
×
1 or
×
0.1 can be selected.
Note 1: Ramp rate can be changed during execution of ramp control. When changed, ramp control is carried out immediately at the
new ramp rate.
How to Execute Ramp Control
Conditions for execution (common to front key input and external input)
(1) AT is not in execution.
(2) STANBY is not on.
(3) RAMP_Down or RAMP_Up is not OFF.
(4) Neither Mode 0 nor Mode 1 not selected in 6-1. 4) Setting of mode.
Note 1: Ramp control is not executed by moving from local SV No. to remote SV No.
Note 2: When YES has been selected in 6-7. 8) Setting of remote tracking, ramp control is not executed by moving from remote SV No.
to local SV No.
Note 3: If power turns OFF during execution of ramp control, ramp control is stopped and the SV value of targeted SV No. serves as SV
value.
To begin with, press the key to call the initial screen (Group 0-0) of the monitor group. Then, pressing the key causes SV No.
to flash. When target SV No. is selected by means of the or key and is registered by means of the key, the RUN
monitor lamp begins to flash and ramp control is executed.
Note 4: A change of mode in 6-1. 4) Setting of mode is not possible during execution of ramp control.
* For temporarily stopping ramp control which is being executed, see 6-8. 5) Temporary stop/restart of ramp control.
- 36 -
RAMP Down: OFF
RAMP Up : OFF
RAMP Unit:
°
C/Min
RAMP Rate: X1
Setting range: Unit/Sec., Unit/Min.
Initial value: Unit/Min.
RAMP Down: OFF
RAMP Up : OFF
RAMP Unit:
°
C/Min
RAMP Rate: X1
Setting range:
×
1,
×
0.1
Initial value:
×
1
D I S P
E N T
Ramp Down: 0.01
°
C/S
Ramp Up : 0.02
°
C/S
Ramp Unit:
°
C/Min
Ramp Rate: X0.1
E N T
DISP
SVNo.
02
DEV
0.886
25.000
°
C
-5
0
+5
SVNo.
02
DEV
0.886
25.000
°
C
-5
0
+5
SVNo.
01
DEV
0.386
30.000
°
C
-5
0
+5
SVNo.
01
DEV
0.886
30.000
°
C
-5
0
+5