Chapter 5 Test and Material List
the TEMP parameter should be 35±0.1 , 37±0.1 and 39±0.1 .
℃
℃
℃
2. CY-F1 probe
Select CY-F1 probe from the manufacturer menu, select CY-F1 temperature
probe as the test fixture, set the analog resistance to 6.534K, 6.018K and
5.548K. Then the TEMP parameter should be 35±0.1 , 37±0.1 and
℃
℃
39±0.1 .
℃
5.1.5 NIBP
Connect the NIBP simulator, adult cuff and accessories, and then connect the module
CUFF and clockwise screw it tightly.
1.
After the simulator self-test, press <ENT> to enter the ADULT analog blood
pressure mode. Set the blood pressure to the 255/195/215 mmHg level, SHIFT
to +15, and the HR to 80BPM. Set PM-9000 Express to the adult mode. Press
<START>. Then the results will be obtained in about 30s. The measured results
should be respectively 270±8mmHg, 210±8mmHg and 230±8mmHg.
2.
Press <ESC> and <
↓
> on the simulator to enter the NEONATE mode. Set the
blood pressure to the 120/80/90 mmHg level, HR to 120bmp, and PM-9000
Express to the pediatric mode. Press <START>. Then the results will be
obtained in about 30s. The measured results should be respectively
120±8mmHg, 80±8mmHg and 90±8mmHg.
3.
Press <ESC> and <
↓
> on the simulator to enter the NEONATE mode. Set the
blood pressure to the 60/30/40 mmHg level, SHIFT to -20, HR to 120bmp, and
PM-9000 Express to the neonate mode. Change the simulator accessory to the
neonatal cuff. Press <START>. Then the results will be obtained in about 30s.
The measured results should be respectively 40±8mmHg, 10±8mmHg and
20±8mmHg.
5.1.6 SpO
2
Select PLETH as the HR source of PM-9000 Express, and put the finger into the SpO
2
sensor. The screen should display the PR and SpO
2
values normally. The normal SpO
2
value is above 97%.
5.1.7 IBP
1. Test fixture
Physiological signal simulator
2. Test procedure
IBP1 test:
①
Set the BP sensitivity of the ECG simulator to 5uv/v/mmHg, BP to 0mmHG, and the
IBP channel 1 to ART. Enter the IBP PRESSURE ZERO menu of the PM-9000
Express, zero Channel 1, and then return to the main screen. Set the BP of the
simulator to 200mmHg. Enter the IBP PRESSURE CALIBRATE menu of the
PM-9000 Express, conduct calibration, and then exit the IBP PRESSURE
CALIBRATE menu.
Set the BP value of the simulator respectively to 40mmHg, 100mmHg and
200mmHg. Then the screen of the PM-9000 Express should display 40±1mmHg,
100±2mmHg and 200±4mmHg.