Chapter 3 Product Specifications
3.15 CO
2
Specifications
Measurement technique
Infrared absorption technique
Measurement mode
Sidestream, microstream or mainstream (optional)
Displayed parameter
EtCO2, FiCO2, Respiration Rate
CO2 function
Meet the requirements of EN864 and ISO9918.
Mindray CO
2
Specifications
CO2 measurement range
0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2mmHg
±5%
±10%
Deflation rate
100, 150ml/min
Precision of deflation rate
15%
Start-up time of CO2
module
< 1min.
The module enters the warming up status after the startup. Ten
minutes later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time
< 240 ms (10 to 90%)
Delay time
< 2s (Length of sampling tube: 7 feet; inner diameter: 0.055
inches; sampling flow: 150ml/min)
Apnea alarm delay
AwRR: 10 to 40 s