Shenchen Peristaltic Pump
V6-12L Series
- 19 -
Internal isolation 5VDC output
②
. External control start/stop, cw/ccw signal input terminal:
Active signal input.
If need change to 12VDC or 24VDC input, please open the controller
housing, and change the jumper connection on the external control
board as below :
GD2:
External control signal common input terminal.
NC:
Null.
CW/CCW:
External control direction signal input
Pulse mode
: the direction change once when receive one
pulse signal
(
rising edge effectively
)
.
Level mode
: when high level the pump clockwise running,
when low level, the pump counterclockwise running.
R/S 2:
External control start/stop signal input
Pulse mode:
the pump working status change once when
receive one pulse signa
(
rising edge effectively
)
l.
Level mode:
when high level, the pump running; when low
level the pump stop running.
Set up the external control mode in the setting interface, turn on
the correspond external control function, external control signal is