Shenchen Pump LabV Series
- 16 -
3.
Dispensing
Boot into
the main
interface
Click the setting button,
choose pump head and
tubing size, choose
working mode: Flow rate
Click the Calibration
button, calibrate the set up
volume, the process refer
to the calibration interface
instruction
Back to the main inter-
face after finish
calibration, click start
key start working
The pump display the current
volume and countdown time,
when the time is out, the pump
stop working automatically,
click the start key, start next
dispensing
Click Dispensing button,
turn on the Dispensing
function, set the
parameters
The flow rate can be
adjust during the pump
working, improve the
accuracy
Display the
dynamic
working state
Display the
current rotating
speed and state
The set up volume
The current
countdown time
The volume has
been transferred
Note: Flow rate calibration process please refer the flow rate calibration
interface instruction.
4.
Timer Start And Stop Function
Boot into
the main
interface
Click the setting
button, choose pump
head and tubing size,
choose working mode:
Flow rate mode
During the pump working,
you also can set the start and
stop time. When fixed
volume or dispensing
function is turn on, the timer
stop function is forbidden
Click Timing button,
set the start time and
over time, set the
repeat times
On the top right corner of the
main interface, means the timer
start and stop function is turn
on, when reach the set up time,
the pump will start or stop
automatically
The timer start and stop
function is turn on
Under the flow rate transferring mode, set the pump start at 8:30 a.m. from
Monday to Friday, stop at 5:30 p.m., the process as below: