Shenchen Pump LabV Series
- 15 -
Boot into
the main
interface
If fixed volume
or dispensing
function turn on
Turn off fixed
volume or disp-
ensing function
YES
Click the setting button,
choose pump head and
tubing size, choose
working mode: Flow rate
Click the flow rate
display area,
input flow rate
Click the Calibra-
tion button, cali-
brate the flow rate
Back to the main inter-
face after finish the
calibration, click the
start key
During the pump working,
you can enter the calibr-
ation interface to micro
adjust the flow rate, impr-
ove the flow rate accuracy
Click here input flow rate
Display the
rotating speed
with thecurrent
flow rate
Display the
running state
and direction
NO
Note: Flow rate calibration process please refer the flow rate calibration
interface instruction.
2.
Fixed Volume Measurement Function
Boot into the
main interface
Click the setting button,
choose pump head and
tubing size, choose
working mode: Flow rate
Back to the main
interface after finish
the calibration, click
the start key
The pump display the current flow
volume and running time, when the
flow volume reach set up value, the
pump will stop automatically. Click
the start key, start the volume
measurement again
Click the Volume button,
turn on the fixed volume
function, the volume range
is 0.01mL-9999L
The flow rate can be
changed during the pump
working, the measurement
accuracy will be lower
after change the flow rate
Click the flow rate
display area in main
interface, input flow
rate
Click the Calibration
button,the default
calibration time is 60s,
you can change it
before calibration
Display the
dynamic
working state
Display the
current rotating
speed and state
The set up volume
The current
running time
The volume has
been transferred
Click here input flow rate
Note: Flow rate calibration process please refer the flow rate calibration
interface instruction.