background image

 

10

 

Microprocessor Control Water Baths 

 

1.  Place  a  calibrated  reference  thermometer  in  the  bath  (a  thermometer  clip  is  provided  with  the 

accessories package). (See Figure 2 for placement.) Allow the thermometer to reach temperature and 
remain stable for one (1) hour.  

2.  Compare the reading on the reference thermometer with the temperature control display.  If there is a 

difference, put  the  display  into  calibrate  mode  by  pressing  both the Up and Down arrow pads at the 
same time until the two (2) outside decimal points begin to blink.   

3.  When the decimal points are blinking, press the Up or Down arrow pad to adjust the display to match 

the reference thermometer.  If the arrow pads are not pressed within five (5) seconds the display will 
revert to showing the temperature within the bath.   

4.  Allow the unit to stabilize again, and repeat calibration if necessary. 

5.  On the W614M, both tanks will need to be calibrated individually. 

Setting the Overtemperature Limit Thermostat 

To set the Over Temperature Limit Thermostat, perform the following steps: 

1.  Verify that the Limit Thermostat was set to its maximum position to allow the water bath to stabilize. 

2.  Turn the Limit Thermostat counterclockwise until the Safety indicator light turns on. 

3.  Turn the Limit Thermostat clockwise until the OTL indicator light turns off. 

4.  Turn the thermostat clockwise again, two (2) of the smallest divisions on its scale past the point where 

the indicator light went out.  This will set the Safety Thermostat at approximately 1

C above the Main 

Temperature set point. 

Figure 2. Thermometer Placement 

 

9850563 

 

 

Note: 

Water bath thermometer must be placed so that the bulb is always fully immersed. The thermometer 
must not touch the sides or the bottom of the bath tank.  

Bath Cover Assembly  

To assemble the bath cover, perform the following steps: 

1.  Insert two (2) self-tapping screws through the respective cover holes and into the holes in the handle, 

leaving the first self-tapping screw loose to assure proper alignment. See Figure 3. Firmly fasten both 
screws into place. Do not over-tighten.  

Figure 3.  Bath Cover Assembly

 

 

Handle Part #3800609

 

Содержание SWB15

Страница 1: ...2 SWB15 SWB15 2 SWB23 SWB23 2 SWB715 SWB 715 2 Previously known as W2M W2M 2 W6M W6M 2 W14M W14M 2 W20M W20M 2 W614M W614M 2 115v 220v and PC Models MICROPROCESSOR CONTROLLED INSTALLATION AND OPERATIO...

Страница 2: ...ile or combustible materials are being heated These units have been tested to the following requirements CAN CSA C22 2 No 61010 1 2012 CAN CSA C22 2 No 61010 2 010 R 2009 UL 61010 1 2004 R 2005 07 R 2...

Страница 3: ...amination 7 CONTROLS OVERVIEW 8 Power Switch 8 Main Temperature Controller 8 Fuses 8 Over Temperature Limit Thermostat OTL 8 Heating Activated 8 Over Temperature Activated Light 8 OPERATION 9 Turning...

Страница 4: ...nual before using the water bath Be sure you follow any city county or other ordinances in your area regarding the use of this unit Use only approved accessories Do not modify system components Any al...

Страница 5: ...delivered the unit If the carton is not damaged open the carton and remove its contents Verify that all of the following equipment is included in the crate One 1 water bath One 1 bath cover two 2 for...

Страница 6: ...p re secteur AC I Indicates the power is ON Rep re de la position MARCHE de l interrupteur d alimentation O Indicates the power is OFF Rep re de la position ARR T de l interrupteur d alimentation Indi...

Страница 7: ...ts are intended for 50 60 HZ application A separate circuit is recommended to prevent damage to the unit due to overloading or circuit failure Location In selecting a location consider all conditions...

Страница 8: ...uit breaker and will cut off power to the unit if there is an electrical surge or malfunction The cause should be determined and corrected before placing the fuse Fuses must be in place for the unit t...

Страница 9: ...ll be necessary to check the water level more frequently 4 Push the Main power switch to the ON position and turn the Over Temperature Limit Thermostat to its maximum position clockwise Setting Main T...

Страница 10: ...Thermostat To set the Over Temperature Limit Thermostat perform the following steps 1 Verify that the Limit Thermostat was set to its maximum position to allow the water bath to stabilize 2 Turn the...

Страница 11: ...ak through openings and cracks and get on electrical components or that may contain solvents corrosives or abrasives that will harm the stainless steel coatings 2 If a hazardous material substance has...

Страница 12: ...g temperatures or in applications where a cover is not used more frequent water level checking is required Test tube Racks If the water bath tank boils dry while containing plastic ware the plastic wi...

Страница 13: ...n Temperature too high Insufficient quantity of water Fill bath a minimum of three fourths 3 4 full with water Main controller set too high See Setting Main Temperature Control Main controller failed...

Страница 14: ...aintain set point Insufficient quantity of water Fill bath a minimum of three fourths 3 4 full with water Temperature set too low Assure that set point is at least 5 degrees over ambient room temperat...

Страница 15: ...B7 2 W6M 2 9751191 9751191 Gable Covers SWB15 W14M SWB15 2 W14M 2 9751184 9751184 Gable Covers SWB23 W20M SWB23 2 W20M 2 9751185 9751185 Cable Covers SWB715 W614M SWB715 2 W614M 2 9751191 9751184 9751...

Страница 16: ...13 Tank Gasket 3450538 3450538 Tanks SWB2 W2M SWB2 2 W2M 2 7930521 7930521 Tanks SWB7 W6M SWB7 2 W6M 2 SWB715 W614M SWB715 2 W614M 2 7930523 7930523 Tanks SWB15 W14M SWB15 2 W14M 2 SWB715 W614M SWB715...

Страница 17: ...1 in 37 47 x 31 12 x 30 73 cm 11 8 x 5 8 x 2 5 in 29 98 x 14 73 x 6 35 cm SWB7 SWB72 2 W6M W6M 2 14 75 x 12 25 x 12 1 in 37 47 x 31 12 x 30 73 cm 11 8 x 5 8 x 6 in 29 98 x 14 73 x 15 24 cm SWB15 SWB1...

Страница 18: ...02 W20M W20M 2 14 17 SWB715 SWB715 2 W614M W614M 2 4 7 small tank 11 225 large tank 5 75 small tank 14 75 large tank Model Range Uniformity Sensitivity SWB2 SWB2 2 W2M W2M 2 Amb 5 C to 80 C 2 C 37 C 1...

Страница 19: ...240 V 2 A 50 60 Hz SWB7 SWB72 2 W6M W6M 2 Volts 100 120 V 3 A 50 60 Hz Volts 220 240 V 2 A 50 60 Hz SWB15 SWB15 2 W14M W14M 2 Volts 100 120 V 5 A 50 60 Hz Volts 220 240 V 3 A 50 60 Hz SWB23 SWB2302 W2...

Страница 20: ...505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750960 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 41 8 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL...

Страница 21: ...4 TEMPERATURE CONTROL 1750961 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 153 6 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK 2800502 EMI FI...

Страница 22: ...302 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750957 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 41 8 BLACK CHERRY ROUND...

Страница 23: ...NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750958 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 153 6 BLACK CHERRY ROUND POWER...

Страница 24: ...1304 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750955 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUN...

Страница 25: ...T NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750956 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER...

Страница 26: ...1306 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750953 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUN...

Страница 27: ...T NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750954 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER...

Страница 28: ...LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 POWER LIGHT 4650554 TEMPERATURE C...

Страница 29: ...4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK 2800502 EMI FILTER HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 POWER LIGHT 4650554 T...

Отзывы: