USER MAINTENANCE
Continued on next page
Calibration Continued
3
Place the controller display in the Operations menu.
a.
Press and hold both the
Up
and
Down
Arrow buttons
simultaneously for approximately 3 seconds.
b.
Release the buttons when “A1” appears on the top display
line and “oPEr” appears in the mid display line.
Operations Menu
4
Advance through the Operations menu to the Temperature
Calibration offset parameter.
a.
Push the green
Advance button
repeatedly unit “i.CA”
appears on the green mid display line and a number value
on the red top line.
5
Adjust the number value in the top display to match the offset
value from step 2a, using the arrow buttons.
6
Save the calibration offset and return to the homepage.
a.
Push the Reset button repeatedly until the display shows
the homepage.
•
The oven will begin to heat or passively cool to reach the
current set point using the offset display value for the
current temperature.
7
Allow the oven to stabilize after achieving the temperature set
point using the offset display value.
Note:
The unit is stabilized when no fluctuations are detected of
±0.2°C
or
greater
are detected for 30 minutes.